AI Article Synopsis

  • The study aimed to evaluate the implantation of a stented decellularised tissue-engineered heart valve (sdTEHV) in sheep to assess its functionality and the tissue's ability to repopulate.
  • Prototypes were successfully implanted in 15 sheep, with regular monitoring showing normal pressure gradients, though some issues with regurgitation were noted due to design flaws.
  • The findings indicate that sdTEHV can be safely used and shows promise for tissue self-maintenance, with further improvements needed in valve design to prevent regurgitation.

Article Abstract

Aims: The objective was to implant a stented decellularised tissue-engineered heart valve (sdTEHV) percutaneously in an animal model, to assess its in vivo functionality and to examine the repopulation and remodelling of the valvular matrix by the recipient's autologous cells.

Methods And Results: Prototypes of sdTEHV were cultured in vitro, decellularised and percutaneously implanted into the pulmonary position in 15 sheep. Functionality was assessed monthly by intracardiac echocardiography (ICE). Valves were explanted after eight, 16 or 24 weeks and analysed macroscopically, histologically and by electron microscopy. Implantation was successful in all animals. Valves showed normal pressure gradients throughout the study. Due to a suboptimal design with small coaptation area, stent ovality led to immediate regurgitation which continuously increased during follow-up. Analyses revealed complete endothelialisation and rapid cellular repopulation and remodelling of the entire matrix. Valves were free from endocarditis, calcification and graft rejection.

Conclusions: sdTEHV can be safely implanted percutaneously. The fast autologous recellularisation and the extensive matrix remodelling demonstrate the valve's potential as a next-generation percutaneous prosthesis with the capacity for tissue self-maintenance and longevity. Regurgitation may be prevented by valve design optimisation.

Download full-text PDF

Source
http://dx.doi.org/10.4244/EIJV12I1A12DOI Listing

Publication Analysis

Top Keywords

vivo functionality
8
matrix remodelling
8
repopulation remodelling
8
percutaneous pulmonary
4
pulmonary valve
4
valve replacement
4
replacement completely
4
completely tissue-engineered
4
tissue-engineered off-the-shelf
4
off-the-shelf heart
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!