Background: We developed a data-driven respiratory motion (RM) correction method (REGAT program) for multiple-pinhole detector CZT SPECT. We verified its clinical feasibility with myocardial perfusion imaging (MPI) and studied its impact on image characteristics.

Methods: This retrospective study included 18 patients having stress/rest Tc-Tetrofosmin MPI SPECT. List mode was acquired on CZT SPECT and processed with REGAT. REGAT generates reconstructed RM-gated volumes that are summed either without realignment (NR-SPECT) or after realignment (R-SPECT). For both stress and rest, we calculated the maximal RM in the 3 axis, and image characteristics of both R-SPECT and NR-SPECT: minimum left ventricular (LV) cavity counts (LV-Min), maximum LV myocardial counts (LV-Max), LV contrast, and FWHM of both anterior (FWHM-ant) and inferior (FWHM-inf) LV myocardial walls.

Results: At both stress and rest, cranio-caudal motion was the dominant axial movement and REGAT had a positive impact on image characteristics as reflected by variations between R-SPECT and NR-SPECT in LV-Min, LV-Max, FWHM-ant, FWHM-inf, and contrast. These latter were well correlated to the amplitude of cranio-caudal motion at both stress and rest.

Conclusions: Data-driven RM correction of MPI acquired with CZT SPECT is clinically feasible and easily applicable. It presents interesting impact on image characteristics.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s12350-016-0493-4DOI Listing

Publication Analysis

Top Keywords

czt spect
16
impact image
12
image characteristics
12
respiratory motion
8
motion correction
8
myocardial perfusion
8
acquired czt
8
stress rest
8
r-spect nr-spect
8
cranio-caudal motion
8

Similar Publications

Article Synopsis
  • This study examined the effectiveness of bone SPECT imaging with and without SwiftScan technology in patients with prostate cancer bone metastases.
  • Involving 10 patients and 130 metastatic lesions, the research compared the quality of images produced at various acquisition times, showcasing improved quality scores and correlation coefficients with SwiftScan.
  • Findings indicated that SwiftScan allowed for a 50% reduction in acquisition time while still maintaining acceptable image quality and accurate quantitative measurements like SUV and contrast-to-noise ratios.
View Article and Find Full Text PDF

Aim: Coronary artery disease (CAD) is a primary cause of mortality, prompting ongoing research into noninvasive diagnostic modalities. This study aimed to evaluate the diagnostic efficacy of exercise electrocardiography testing (EET) combined with cadmium zinc tellurium cardiac-dedicated single photon emission computed tomography (CZT-SPECT) imaging for CAD.

Materials And Methods: CZT-SPECT and EET were examined in 124 patients aged 20-85 years, followed by coronary angiography to evaluate the sensitivity, specificity, positive predictive value, and negative predictive value of EET/CZT-SPECT alone and in combination.

View Article and Find Full Text PDF

This study aimed to assess the diagnostic capability of quantitative parameters from whole-body bone single-photon emission computed tomography/computed tomography (SPECT/CT) in detecting bone metastases in prostate cancer patients; Methods: We retrospectively analyzed 82 prostate cancer patients who underwent staging bone scintigraphy with a full-ring 360° Cadmium-Zinc-Telluride (CZT) SPECT/CT system. From the SPECT/CT images, we measured the maximum (SUVmax) and mean (SUVmean) standardized uptake values at six normal bone sites (skull, humerus, thoracic spine, lumbar spine, iliac bone, and femur), and the SUVmax for both metastatic and benign bone lesions. Ratios of lesion SUVmax-to-maximum and mean uptake values at the skull, humerus, and femur were computed for each lesion; Results: SUVmax and SUVmean at the skull and femur exhibited significantly lower variance compared to those at the thoracic spine, lumbar spine, and iliac bone, and revealed no significant differences between patients with and without bone metastasis.

View Article and Find Full Text PDF

Recent Progress in Myocardial Perfusion Imaging Techniques.

Nucl Med Mol Imaging

December 2024

Department of Nuclear Medicine, Severance Hospital, Yonsei University College of Medicine, Seoul, Korea.

Article Synopsis
  • Nuclear cardiology has advanced rapidly, particularly in myocardial perfusion imaging (MPI) since the 1970s, which is vital for assessing and managing coronary artery disease.
  • Recent innovations include the development of gamma cameras with cadmium zinc telluride (CZT) systems and enhancements in software, as well as ongoing studies into new radiopharmaceuticals for positron emission tomography (PET)-MPI.
  • Despite these advancements, challenges like high equipment and examination costs, along with limited accessibility, pose significant barriers to the practical use of these technologies in nuclear medicine.
View Article and Find Full Text PDF

New Generation SPECT Cameras Based on Cadmium-Zinc Telluriide Technology.

Ann Nucl Cardiol

October 2024

Division of Nuclear Medicine, Department of Radiology, Saitama Medical University Hospital, Saitama, Japan.

Article Synopsis
  • Despite advances in positron emission tomography, cardiac SPECT imaging remains popular due to its accessibility and widespread clinical application.
  • CZT-based detectors offer improved imaging capabilities compared to traditional Anger type detectors, leading to successful clinical use since their introduction 15 years ago.
  • Ongoing research into new semiconductor materials and the development of advanced SPECT systems indicate potential future advancements in nuclear cardiology.
View Article and Find Full Text PDF