Purpose: To compare the impact of a 100 kV tube voltage protocol to 120 kV in terms of image quality and radiation dose by a 320 row coronary computed tomography angiography (CCTA) with automatic exposure control (AEC).

Materials And Methods: Using a propensity matched analysis we compared a group of 135 patients scanned using a 100 kV tube voltage protocol with a group of 135 subjects scanned employing a 120 kV tube voltage setting. In all subjects the heart rate (HR) was <65 bpm and all CT scans were acquired using a prospective ECG gating and AEC strategy. Mean effective radiation dose and subjective and objective (Noise or N, signal to noise ratio or SNR, contrast to noise ratio or CNR) image quality, were evaluated. Subjective quality was assessed by two experienced radiologists using a 5-point scale (0: non diagnostic-4: excellent) using the 15-segment American Heart Association (AHA) coronary artery classification.

Results: Mean effective dose and noise were non significantly different between the two groups: mean effective dose was 2.89 ± 0.7 mSv in the 100 kV group and 2.80 ± 0.57 mSv in the 120 kV group (p = 0.25) while noise was 28.9 ± 3.3 in the 120 kV group and 29.05 ± 3.6 in the 100 kV group (p = 0.72). Both SNR and CNR were significantly higher in the 100 kV group than in the 120 kV group. This data agrees with the evidence that subjective quality was significantly higher in the 100 kV group in the middle and distal segmental classes.

Conclusion: Our study shows that, in using a 320 row CCTA with AEC strategy it is better to employ a 100 kV tube voltage protocol because compared to 120 kV tube voltage setting, it appears to significantly improve both subjective and objective image quality without decreasing the mean effective radiation dose.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s11547-016-0643-4DOI Listing

Publication Analysis

Top Keywords

tube voltage
12
coronary computed
8
computed tomography
8
tomography angiography
8
angiography ccta
8
ccta automatic
8
automatic exposure
8
exposure control
8
image quality
8
100 kv tube
8

Similar Publications

Dynamic Digital Radiography (DDR) in the Diagnosis of a Diaphragm Dysfunction.

Diagnostics (Basel)

December 2024

Postgraduation School in Radiodiagnostics, Università degli Studi di Milano, Via Festa del Perdono 7, 20122 Milan, Italy.

Dynamic digital radiography (DDR) is a recent imaging technique that allows for real-time visualization of thoracic and pulmonary movement in synchronization with the breathing cycle, providing useful clinical information. A 46-year-old male, a former smoker, was evaluated for unexplained dyspnea and reduced exercise tolerance. His medical history included a SARS-CoV-2 infection in 2021.

View Article and Find Full Text PDF

Polymer nanocomposites have been investigated as lightweight and suitable alternatives to lead-based clothing. The present study aims to fabricate flexible, lead-free, X-ray-shielding composites using a polyvinyl chloride (PVC) matrix and different nanostructures. Four different nanostructures containing impure tungsten oxide, tungsten oxide (WO), barium tungstate (BaWO), and bismuth tungstate (BiWO) were synthesized through various methods.

View Article and Find Full Text PDF

Purpose: With the widespread introduction of dual energy computed tomography (DECT), applications utilizing the spectral information to perform material decomposition became available. Among these, a popular application is to decompose contrast-enhanced CT images into virtual non-contrast (VNC) or virtual non-iodine images and into iodine maps. In 2021, photon-counting CT (PCCT) was introduced, which is another spectral CT modality.

View Article and Find Full Text PDF

Establishment of Local Diagnostic Reference Levels for Head CT Imaging in the Madina Region, Saudi Arabia.

Diagnostics (Basel)

December 2024

Medical Diagnostic Imaging Department, College of Health Science, University of Sharjah, Sharjah 27272, United Arab Emirates.

: Computed Tomography (CT) is crucial in medical diagnosis, particularly for head examinations. Diagnostic Reference Levels (DRLs) are pivotal in balancing diagnostic efficacy with radiation safety. International organizations such as the International Atomic Energy Agency (IAEA) and the International Commission on Radiological Protection (ICRP) provide guidelines for establishing DRLs, emphasizing their importance in optimizing radiation doses.

View Article and Find Full Text PDF
Article Synopsis
  • The study investigated the noise reduction capabilities of a photon-counting detector (PCD) in computed tomography (CT) using a model-based iterative reconstruction algorithm (QIR).
  • Forty repeated scans were conducted on a water phantom and compared with a conventional energy-integrating detector (EID) to assess noise characteristics.
  • Results showed that PCD-CT significantly reduced noise levels and improved image uniformity, demonstrating the effectiveness of QIR in decreasing noise without altering the overall distribution of noise values.
View Article and Find Full Text PDF