Improving Displacement Signal-to-Noise Ratio for Low-Signal Radiation Force Elasticity Imaging Using Bayesian Techniques.

Ultrasound Med Biol

Department of Biomedical Engineering, Vanderbilt University, Nashville, TN, USA. Electronic address:

Published: August 2016

AI Article Synopsis

  • Radiation force-based elasticity imaging is being researched for clinical applications like liver fibrosis staging and cardiovascular tissue characterization.
  • The study assesses how peak displacement magnitude impacts image quality and suggests using a Bayesian estimator to improve data collection in low-signal conditions.
  • Results indicate that the Bayesian estimator enhances signal-to-noise ratio (SNR) by about 9 dB and doubles the contrast-to-noise ratio, suggesting it can effectively boost data quality in challenging imaging scenarios.

Article Abstract

Radiation force-based elasticity imaging is currently being investigated as a possible diagnostic modality for a number of clinical tasks, including liver fibrosis staging and the characterization of cardiovascular tissue. In this study, we evaluate the relationship between peak displacement magnitude and image quality and propose using a Bayesian estimator to overcome the challenge of obtaining viable data in low displacement signal environments. Displacement data quality were quantified for two common radiation force-based applications, acoustic radiation force impulse imaging, which measures the displacement within the region of excitation, and shear wave elasticity imaging, which measures displacements outside the region of excitation. Performance as a function of peak displacement magnitude for acoustic radiation force impulse imaging was assessed in simulations and lesion phantoms by quantifying signal-to-noise ratio (SNR) and contrast-to-noise ratio for varying peak displacement magnitudes. Overall performance for shear wave elasticity imaging was assessed in ex vivo chicken breast samples by measuring the displacement SNR as a function of distance from the excitation source. The results show that for any given displacement magnitude level, the Bayesian estimator can increase the SNR by approximately 9 dB over normalized cross-correlation and the contrast-to-noise ratio by a factor of two. We conclude from the results that a Bayesian estimator may be useful for increasing data quality in SNR-limited imaging environments.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5388359PMC
http://dx.doi.org/10.1016/j.ultrasmedbio.2016.03.004DOI Listing

Publication Analysis

Top Keywords

elasticity imaging
16
radiation force
12
peak displacement
12
displacement magnitude
12
bayesian estimator
12
signal-to-noise ratio
8
radiation force-based
8
displacement
8
data quality
8
acoustic radiation
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!