A hydrogel based rapid test method for detection of Escherichia coli (E. coli) in contaminated water samples.

Analyst

Micro & Nano-scale Transport Laboratory, Lassonde School of Engineering, York University, Toronto, Ontario, Canada.

Published: May 2016

AI Article Synopsis

Article Abstract

We have formulated a new chemical composition for rapid detection of Escherichia coli (E. coli) with currently available enzymatic substrates. We have evaluated the performance of the new chemical composition with different kinds of bacteria, and metallic and ionic interferences and optimized the chemical composition for rapid and specific detection of E. coli. We used a novel hydrogel based porous matrix to encapsulate the optimized chemical compounds and incorporated it within a readily available plunger-tube assembly. This overall system allows efficient, field deployable, rapid testing of water samples by simultaneously pre-concentrating and detecting E. coli within one integrated unit. We were able to detect E. coli concentrations of 4 × 10(6) CFU mL(-1) to 4 × 10(5) CFU mL(-1) within 5 min and 4 × 10(4) CFU mL(-1) to 400 CFU mL(-1) within 60 min using the integrated plunger-tube assembly containing the hydrogel matrix.

Download full-text PDF

Source
http://dx.doi.org/10.1039/c6an00400hDOI Listing

Publication Analysis

Top Keywords

cfu ml-1
16
chemical composition
12
hydrogel based
8
detection escherichia
8
escherichia coli
8
coli coli
8
water samples
8
composition rapid
8
optimized chemical
8
plunger-tube assembly
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!