Mechanical circulatory support offers an alternative therapeutic treatment for patients with dysfunctional single ventricle physiology. An intravascular axial flow pump is being developed as a cavopulmonary assist device for these patients. This study details the development of a new rotating impeller geometry. We examined the performance of 8 impeller geometries with blade stagger or twist angles varying from 100° to 800° using computational methods. A refined range of blade twist angles between 300° and 400° was then identified, and 4 additional geometries were evaluated. Generally, the impeller designs produced 4-26mmHg for flow rates of 1-4L/min for 6000-8000 RPM. A data regression analysis was completed and found the impeller with 400° of blade twist to be the superior performer. A hydraulic test was conducted on a prototype of the 400° impeller, which generated measurable pressure rises of 7-28mmHg for flow rates of 1-4L/min at 6000-8000 RPM. The findings of the numerical model and experiment were in reasonable agreement within approximately 20%. These results support the continued development of an axial-flow, mechanical cavopulmonary assist device as a new clinical therapeutic option for Fontan patients.