Purpose: The objective of this study was to assess the biomechanical properties of bicortical locking screws, unicortical locking screws, and unicortical far-cortex-abutting locking screw fixation in a cadaver model of comminuted midshaft clavicle fractures stabilized with a locking plate placed on the superior surface of the clavicle.

Methods: Nine pairs of adult fresh-frozen cadaver clavicles were allocated into 3 groups for either bicortical, unicortical, or unicortical far-cortex-abutting locking plate fixation. After a 1-cm osteotomy to simulate a comminuted fracture and instrumentation with an 8-hole locking plate placed on the superior surface of the clavicle, the specimens were mounted in a custom dual-gimbal fixture in a materials-testing system and tested in axial compression, torsional, and torsional load to failure.

Results: Axial stiffness and axial osteotomy site stiffness did not demonstrate differences between constructs. In cyclical torsion, both the bicortical and the unicortical far-cortex-abutting constructs were significantly stiffer than the unicortical construct. For torsional failure stiffness, both the bicortical and the unicortical far-cortex-abutting constructs were significantly stiffer than the unicortical construct. There was no difference between bicortical and unicortical far-cortex-abutting for torsional failure stiffness. The bicortical construct exhibited significantly higher peak failure torque compared with the unicortical construct.

Conclusions: Unicortical far-cortex-abutting locking screw fixation provides comparable mechanical properties under axial and torsional loads to bicortical fixation, without penetrating the far cortex.

Clinical Relevance: Unicortical far-cortex-abutting locking screw fixation obviates far cortex penetration, and thereby protects nearby anatomical structures, may ease symptomatic implant removal, alleviates refracture risk, and eases conversion to bicortical fixation in the case of revision surgery.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jhsa.2016.04.001DOI Listing

Publication Analysis

Top Keywords

unicortical far-cortex-abutting
32
bicortical unicortical
20
far-cortex-abutting locking
16
unicortical
14
locking screw
12
screw fixation
12
locking plate
12
bicortical
9
unicortical unicortical
8
far-cortex-abutting
8

Similar Publications

Purpose: The objective of this study was to assess the biomechanical properties of bicortical locking screws, unicortical locking screws, and unicortical far-cortex-abutting locking screw fixation in a cadaver model of comminuted midshaft clavicle fractures stabilized with a locking plate placed on the superior surface of the clavicle.

Methods: Nine pairs of adult fresh-frozen cadaver clavicles were allocated into 3 groups for either bicortical, unicortical, or unicortical far-cortex-abutting locking plate fixation. After a 1-cm osteotomy to simulate a comminuted fracture and instrumentation with an 8-hole locking plate placed on the superior surface of the clavicle, the specimens were mounted in a custom dual-gimbal fixture in a materials-testing system and tested in axial compression, torsional, and torsional load to failure.

View Article and Find Full Text PDF

Purpose: To provide comparative biomechanical evaluation of bicortical locking versus unicortical-abutting locking screw-plate fixation in a comminuted radius fracture model.

Methods: A validated synthetic substitute of the adult human radius with a 1.5-cm-long segmental mid-diaphyseal defect was used in the study to simulate a comminuted fracture.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!