Objectives: Metal artefacts can impair accurate diagnosis of haemorrhage using flat detector CT (FD-CT), especially after aneurysm coiling. Within this work we evaluate a prototype metal artefact reduction algorithm by comparison of the artefact-reduced and the non-artefact-reduced FD-CT images to pre-treatment FD-CT and multi-slice CT images.

Methods: Twenty-five patients with acute aneurysmal subarachnoid haemorrhage (SAH) were selected retrospectively. FD-CT and multi-slice CT before endovascular treatment as well as FD-CT data sets after treatment were available for all patients. The algorithm was applied to post-treatment FD-CT. The effect of the algorithm was evaluated utilizing the pre-post concordance of a modified Fisher score, a subjective image quality assessment, the range of the Hounsfield units within three ROIs, and the pre-post slice-wise Pearson correlation.

Results: The pre-post concordance of the modified Fisher score, the subjective image quality, and the pre-post correlation of the ranges of the Hounsfield units were significantly higher for artefact-reduced than for non-artefact-reduced images. Within the metal-affected slices, the pre-post slice-wise Pearson correlation coefficient was higher for artefact-reduced than for non-artefact-reduced images.

Conclusion: The overall diagnostic quality of the artefact-reduced images was improved and reached the level of the pre-interventional FD-CT images. The metal-unaffected parts of the image were not modified.

Key Points: • After coiling subarachnoid haemorrhage, metal artefacts seriously reduce FD-CT image quality. • This new metal artefact reduction algorithm is feasible for flat-detector CT. • After coiling, MAR is necessary for diagnostic quality of affected slices. • Slice-wise Pearson correlation is introduced to evaluate improvement of MAR in future studies. • Metal-unaffected parts of image are not modified by this MAR algorithm.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00330-016-4351-1DOI Listing

Publication Analysis

Top Keywords

reduction algorithm
12
subarachnoid haemorrhage
12
artefact-reduced non-artefact-reduced
12
image quality
12
slice-wise pearson
12
algorithm applied
8
flat detector
8
patients acute
8
metal artefacts
8
fd-ct
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!