Electrocorticography (ECoG) is an important area of research for Brain-Computer Interface (BCI) development. ECoG, along with some other biopotentials, has spectral characteristics that can be exploited for more optimal front-end performance than is achievable with conventional techniques. This paper optimizes noise performance of such a system and discusses an equalization technique that reduces the analog-to-digital converter (ADC) dynamic range requirements and eliminates the need for a variable gain amplifier (VGA). We demonstrate a fabricated prototype in 1p9m 65 nm CMOS that takes advantage of the presented findings to achieve high-fidelity, full-spectrum ECoG recording. It requires 1.08 μW over a 150 Hz bandwidth for the entire analog front end and only 7 bits of ADC resolution.