Performance of a completely automated system for monitoring CMV DNA in plasma.

J Clin Virol

Department of Virology, Toulouse University Hospital, Toulouse, France; Department of Physiopathology, Toulouse Purpan, Unité Inserm U563, Toulouse, France.

Published: June 2016

Background: Completely automated systems for monitoring CMV-DNA in plasma samples are now available.

Objectives: Evaluate analytical and clinical performances of the VERIS™/MDx System CMV Assay(®).

Study Design: Analytical performance was assessed using quantified quality controls. Clinical performance was assessed by comparison with the COBAS(®) Ampliprep™/COBAS(®) Taqman CMV test using 169 plasma samples that had tested positive with the in-house technique in whole blood.

Results: The specificity of the VERIS™/MDx System CMV Assay(®) was 99% [CI 95%: 97.7-100]. Intra-assay reproducibilities were 0.03, 0.04, 0.05 and 0.04 log10IU/ml (means 2.78, 3.70, 4.64 and 5.60 log10IU/ml) for expected values of 2.70, 3.70, 4.70 and 5.70 log10IU/ml. The inter-assay reproducibilities were 0.12 and 0.08 (means 6.30 and 2.85 log10IU/ml) for expected values of 6.28 and 2.80 log10IU/ml. The lower limit of detection was 14.6IU/ml, and the assay was linear from 2.34 to 5.58 log10IU/ml. The results for the positive samples were concordant (r=0.71, p<0.0001; slope of Deming regression 0.79 [CI 95%: 0.56-1.57] and y-intercept 0.79 [CI 95%: 0.63-0.95]). The VERIS™/MDx System CMV Assay(®) detected 18 more positive samples than did the COBAS(®) Ampliprep™/COBAS(®) Taqman CMV test and the mean virus load were higher (0.41 log10IU/ml). Patient monitoring on 68 samples collected from 17 immunosuppressed patients showed similar trends between the two assays. As secondary question, virus loads detected by the VERIS™/MDx System CMV Assay(®) were compared to those of the in-house procedure on whole blood. The results were similar between the two assays (-0.09 log10IU/ml) as were the patient monitoring trends.

Conclusion: The performances of the VERIS™/MDx System CMV Assay(®) facilitated its routine use in monitoring CMV-DNA loads in plasma samples.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jcv.2016.03.024DOI Listing

Publication Analysis

Top Keywords

completely automated
8
plasma samples
8
veris™/mdx system
8
system cmv
8
performance assessed
8
log10iu/ml expected
8
expected values
8
log10iu/ml
6
performance completely
4
automated system
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!