Objective: To analyze the rigidity of a platform-type external fixator assembly, according to different numbers of pins on each clamp.

Methods: Computer simulation on a large-sized Cromus dynamic external fixator (Baumer SA) was performed using a finite element method, in accordance with the standard ASTM F1541. The models were generated with approximately 450,000 quadratic tetrahedral elements. Assemblies with two, three and four Schanz pins of 5.5 mm in diameter in each clamp were compared. Every model was subjected to a maximum force of 200 N, divided into 10 sub-steps. For the components, the behavior of the material was assumed to be linear, elastic, isotropic and homogeneous. For each model, the rigidity of the assembly and the Von Mises stress distribution were evaluated.

Results: The rigidity of the system was 307.6 N/mm for two pins, 369.0 N/mm for three and 437.9 N/mm for four.

Conclusion: The results showed that four Schanz pins in each clamp promoted rigidity that was 19% greater than in the configuration with three pins and 42% greater than with two pins. Higher tension occurred in configurations with fewer pins. In the models analyzed, the maximum tension occurred on the surface of the pin, close to the fixation area.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4799440PMC
http://dx.doi.org/10.1016/S2255-4971(15)30017-3DOI Listing

Publication Analysis

Top Keywords

external fixator
12
pins
8
schanz pins
8
tension occurred
8
relationship rigidity
4
rigidity external
4
fixator number
4
number pins
4
pins computer
4
computer analysis
4

Similar Publications

Effectiveness of non-operative methods of treatment of carpal tunnel syndrome: a narrative review.

Wiad Lek

January 2025

DEPARTAMENT OF GENERAL NAD HAND SURGERY, STUDENT'S SCIENTIFIC CIRCLE, POMERANIAN MEDICAL UNIVERSITY, SZCZECIN, POLAND.

Carpal tunnel syndrome (CTS) can be treated with several methods, including surgical and non-surgical techniques. Non-surgical methods include wrist splinting, systemic pharmacotherapy, intracarpal injections of steroids hydrodissection, acupuncture, nerve and tendon mobilization, osteopathy, taping, topical application of ointments, laser, ultrasound and shock-wave therapies. These treatments are generally less effective than surgery, and provide only short-lived effect, but it may be quite sufficient for a certain category of patients, particularly those suffering from mild symptoms.

View Article and Find Full Text PDF

Revision quadriceps tendon repair: A case series and technique guide to a novel repair.

Trauma Case Rep

February 2025

Department of Orthopaedic Surgery and Sports Medicine, University at Buffalo, Jacobs School of Medicine and Biomedical Sciences, Buffalo, NY 14203, United States of America.

Introduction: Revision quadriceps tendon repair is a challenging problem. In this four-case series, novel quadriceps tendon revision resulted in improved range of motion and durable repair for patients with recurrent rupture.

Methods: Our technique includes a combination of a running locked #5 FiberWire or 2 mm SutureTape suture placed through parallel medial, lateral, and central drill holes in the patella with running Krackow-type quadriceps tendon repair medially and laterally resulting in four strands, delivering the vastus medialis and medial quadriceps tendon to an anatomic repair at the superior pole of the patella, with 2 sutures passed centrally and 1 each passed medially and laterally and then tied.

View Article and Find Full Text PDF

Objective: This study aimed to evaluate the clinical and radiologic findings in the treatment of live oligotrophic and pseudoarthrosis nonunions in cats using a linear external fixator.

Methods: The study included 10 cats of various breeds, ages, and sexes that had previously undergone osteosynthesis at another clinic but did not achieve successful results. These cats were diagnosed with nonunion based on clinical and radiographic examinations conducted at the Clinic of the Surgery Department, Siirt University Animal Health Practice and Research Hospital between 2022 and 2023.

View Article and Find Full Text PDF

Total contact casting is the gold standard for plantar foot ulcers but has been questioned in heel pressure ulcers. Current offloading of heel ulcers is typically removable offloading boots. We describe using a modified posterior splint to offload heel ulcers in nonweightbearing patients.

View Article and Find Full Text PDF

Background: Open pelvic fractures are rare but represent a serious clinical problem with high mortality rates. Acute mortality is often associated with hemorrhage, whereas delayed mortality is most often associated with sepsis and multiple organ failure. We report a case of Wang's classification of type II open pelvic ring fracture with hemorrhagic shock and septic shock from gas gangrene.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!