Purpose: Carotid artery diseases due to plaque buildup at the carotid bifurcation are a leading cause of stroke. Accurate plaque quantification and characterization of plaque composition and morphology by magnetic resonance imaging (MRI) is essential to identifying high-risk patients. Difficulties in detecting plaque, which is physically small, and the unique physiological structure of the carotid artery make use of a radio frequency (RF) coil array with high resolution, large longitudinal coverage, and deep penetration ideal for clinical examinations. The goal of this project was to design and fabricate a sensitive RF coil array with sufficient imaging coverage and signal-to-noise ratio (SNR) for carotid artery imaging at 3 T.
Methods: Based on clinical requirements and the anatomical structure of the human carotid artery, an 8-channel carotid coil array was designed and fabricated for 3 T MRI of the carotid artery in humans. The performance of the proposed 8-channel carotid coil array was validated through bench tests and MR imaging experiments on a 3 T whole body MRI scanner. Its performance was also compared experimentally to the performance of a commercial 4-channel phased array carotid coil designed by Machnet BV (Machnet BV coil, Roden, Netherlands).
Results: The 8-channel carotid coil array performed significantly better in imaging the carotid artery than the commercial 4-channel Machnet BV coil in terms of the SNR, coverage, and penetration depth. In parallel imaging, the proposed 8-channel carotid coil array demonstrated a much lower maximum value and average value of the geometry factor in the region of interest. Carotid artery images acquired in vivo using the proposed 8-channel carotid artery coil and the commercial 4-channel Machnet BV coil were also compared, demonstrating the former's potential for clinical diagnosis.
Conclusions: Based on the analyses of phantom and in vivo imaging studies, the proposed 8-channel carotid coil array has the potential for use in clinical diagnosis, performing better in terms of SNR, imaging coverage, and penetration depth than the commercial 4-channel carotid artery coil array at 3 T. In future studies, the proposed 8-channel carotid coil array can also serve as an important part of a large-scale multichannel coil array for imaging the whole carotid artery system, including the extracranial and intracranial arteries.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1118/1.4944500 | DOI Listing |
Magn Reson Med Sci
January 2025
Department of Biomedical Engineering, Gachon University, Seongnam, Gyeonggi, Korea.
Purpose: Hyperthermia is a treatment that applies heat to damage or kill cancer cells and can be also used for drug deliveries. It is important to apply the heat into the specific area in order to target the cancer tissue and avoid damaging healthy tissue. For this reason, the development of heat applicators that have the capability to deliver the heat to the target area is vital.
View Article and Find Full Text PDFFront Oncol
January 2025
Department of Radiology, the Second Affiliated Hospital of Nanjing Medical University, Nanjing, China.
Background: In the realm of breast cancer diagnosis and treatment, accurately discerning molecular subtypes is of paramount importance, especially when aiming to avoid invasive tests. The updated guidelines for diagnosing and treating HER2 positive advanced breast cancer, as presented at the 2021 National Breast Cancer Conference and the Annual Meeting of the Chinese Society of Clinical Oncology, highlight the significance of this approach. A new generation of drug-antibody combinations has emerged, expanding the array of treatment options for HER2 positive advanced breast cancer and significantly improving patient survival rates.
View Article and Find Full Text PDFIEEE J Solid-State Circuits
November 2024
Department of Electrical and Computer Engineering, Rice University, Houston TX, 77005, USA.
Miniature bioelectronic implants promise revolutionary therapies for cardiovascular and neurological disorders. Wireless power transfer (WPT) is a significant method for miniaturization, eliminating the need for bulky batteries in today's devices. Despite successful demonstrations of millimetric battery-free implants in animal models, the robustness and efficiency of WPT are known to degrade significantly under misalignment incurred by body movements, respiration, heart beating, and limited control of implant orientation during surgery.
View Article and Find Full Text PDFMed Phys
January 2025
Department of Electrical and Electronics Engineering, Bilkent University, Ankara, Turkey.
Background: Radiofrequency (RF) transmit arrays play a crucial role in various MRI applications, offering enhanced field control and improved imaging capabilities. Designing and optimizing these arrays, particularly in high-field MRI settings, poses challenges related to coupling, resonance, and construction imperfections. Numerical electromagnetic simulation methods effectively aid in the initial design, but discrepancies between simulated and fabricated arrays often necessitate fine-tuning.
View Article and Find Full Text PDFMultichannel transceiver coil arrays are needed to enable parallel imaging and B1 manipulation in ultrahigh field MR imaging and spectroscopy. However, the design of such transceiver coils and coil arrays often faces technical challenges in achieving the required high operating frequency at the ultrahigh fields and sufficient electromagnetic (EM) decoupling between resonant elements. In this work, we propose a high impedance microstrip transmission line resonator (HIMTL) technique that has unique high frequency capability and adequate EM decoupling without the use of dedicated decoupling circuits.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!