Objectives: Bone-anchored hearing aids (BAHA) occasionally cause soft tissue problems due to abutment. Because Sophono does not have abutment penetrating skin, it is thought that Sophono has no soft tissue problem relating to abutment. On the other hand, transcutaneous device's output is reported to be 10 to 15 dB lower than percutaneous device. Therefore, in this study, Sophono and BAHA were compared to each other from surgical and audiological points of view.

Methods: We retrospectively reviewed the medical records of 9 Sophono patients and 10 BAHA patients. In BAHA cases, single vertical incision without skin thinning technique was done. We compared Sophono to BAHA by operation time, wound healing time, postoperative complications, postoperative hearing gain after switch on, and postoperative air-bone gap.

Results: The mean operation time was 60 minutes for Sophono and 25 minutes for BAHA. The wound healing time was 14 days for Sophono and 28 days for BAHA. No major intraoperative complication was observed. Skin problem was not observed in the 2 devices for the follow-up period. Postoperative hearing gain of bilateral aural atresia patients was 39.4 dB for BAHA (n=4) and 25.5 dB for Sophono (n=5). However, the difference was not statistically significant. In all patients included in this study, the difference of air-bone gap between two groups was 16.6 dB at 0.5 kHz and 18.2 dB at 4 kHz. BAHA was statistically significantly better than Sophono.

Conclusion: Considering the audiologic outcome, BAHA users were thought to have more audiologic benefit than Sophono users. However, Sophono had advantages over BAHA with abutment in cosmetic outcome. Sophono needed no daily skin maintenance and soft tissue complication due to abutment would not happen in Sophono. Therefore, a full explanation about each device is necessary before deciding implantation.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4792243PMC
http://dx.doi.org/10.21053/ceo.2016.9.1.21DOI Listing

Publication Analysis

Top Keywords

sophono
13
soft tissue
12
baha
11
bone-anchored hearing
8
hearing aids
8
sophono baha
8
patients baha
8
operation time
8
wound healing
8
healing time
8

Similar Publications

Objectives Percutaneous and transcutaneous bone conduction hearing implants are used in various otological conditions. This study aimed to evaluate patient satisfaction with Sophono implants and their long-term surgical outcomes. Methods This retrospective study was conducted at Queen Elizabeth Hospital, Birmingham, between October 2022 and May 2023.

View Article and Find Full Text PDF

 Bone conduction implants, responsible for transmitting sound from an external processor to the inner ear, can be divided into active and passive, depending on the vibratory stimulus location. The use of transcutaneous device has increased, given its aesthetic appeal, the complications and limitations of percutaneous devices, and patient's treatment adherence, focusing mainly on efficacy. However, various complications are associated with the use of transcutaneous prosthesis, which can often be serious.

View Article and Find Full Text PDF

Objective: To assess the outcomes of pediatric Osia 2® System placements.

Methods: We performed a retrospective chart review of primary and revision Osia 2® System surgical cases at two tertiary academic children's hospitals. Operative details and post-operative surgical and audiologic outcomes were recorded.

View Article and Find Full Text PDF

Background: The bone conductive implants (BCI) are nowadays a reliable alternative for rehabilitation of specific forms of hearing loss, i.e. conductive, mixed or single sided deafness (SSD).

View Article and Find Full Text PDF

Osseointegrated Auditory Devices-Transcutaneous: Sophono and Baha Attract.

Otolaryngol Clin North Am

April 2019

Otology- Neurotology, Ear Nose and Throat Associates of New York, 35-30 Francis Lewis Boulevard, Auburndale, NY 11358, USA.

Percutaneous osseointegrated bone conduction auditory devices provide excellent auditory rehabilitation. Device-related complications relate to skin abutment interface and cosmetic concerns, resulting in the development of transcutaneous devices. The Sophono and Baha Attract are safe and considered cosmetically superior to the percutaneous Baha Connect and Ponto.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!