Objective: To develop a computational method to detect and quantify burst suppression patterns (BSP) in the EEGs of critical care patients. A multi-center validation study was performed to assess the detection performance of the method.
Methods: The fully automatic method scans the EEG for discontinuous patterns and shows detected BSP and quantitative information on a trending display in real-time. The method is designed to work without setting any patient specific parameters and to be insensitive to EEG artifacts and periodic patterns. For validation a total of 3982 h of EEG from 88 patients were analyzed from three centers. Each EEG was annotated by two reviewers to assess the detection performance and the inter-rater agreement.
Results: Average inter-rater agreement between pairs of reviewers was κ=0.69. On average 22% of the review segments included BSP. An average sensitivity of 90% and a specificity of 84% were measured on the consensus annotations of two reviewers. More than 95% of the periodic patterns in the EEGs were correctly suppressed.
Conclusion: A fully automatic method to detect burst suppression patterns was assessed in a multi-center study. The method showed high sensitivity and specificity.
Significance: Clinically applicable burst suppression detection method validated in a large multi-center study.