This study seeks to identify important factors related to backside wear of tibial inserts in vivo and determine an appropriate wear model for backside wear. An IRB approved database was queried for tibial inserts of a single design from one manufacturer that exhibited evidence of rotatory motion on the backside of the polyethylene. These devices were measured for volumetric wear using a previously established protocol. Features including the change in locking lip width and measurement of micro-motion marks were used to describe the motion pattern. Volumetric wear and implant characteristics were compared using linear regressions by modeling wear theories suggested by Archard and Wang to determine the most appropriate model for backside wear. The Wang model showed that duration, adjusted sliding distance, and cross-shear index accounted for approximately 58% of the volumetric wear variation while adjusted sliding distance and duration in vivo accounted for approximately 35% of the volumetric wear variation in the Archard model. Patient weight (p = 0.750), patient BMI (p = 0.680), and backside area (p = 0.784) of the tibial insert were all found to be non-significant in the Wang model. Similarly, patient weight (p = 0.233), patient BMI (p = 0.162), and backside area (p = 0.796) were found to be non-significant in the Archard model. Multidirectional micro-motion appears to contribute significantly to the wear of these components, supporting the Wang theory of cross-shear for polyethylene wear. Cross-shear of polymers on an unpolished titanium tray can lead to an increase in wear debris in the body. Care should be taken when designing locking mechanisms and tray designs. © 2016 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 34:1933-1940, 2016.

Download full-text PDF

Source
http://dx.doi.org/10.1002/jor.23203DOI Listing

Publication Analysis

Top Keywords

backside wear
16
volumetric wear
16
wear
13
tibial inserts
8
determine appropriate
8
model backside
8
wang model
8
adjusted sliding
8
sliding distance
8
wear variation
8

Similar Publications

Background: The use of uncemented cups during total hip arthroplasty (THA) has gained popularity in recent years. The Robert Mathys (RM) pressfit cup, an uncemented monoblock implant is expected to preserve bone density due to its composition and external surface, while reducing backside wear with its monoblock construction. These factors should lead to a high survival rate of the implant.

View Article and Find Full Text PDF

Background: Highly cross-linked polyethylene (HXLPE) was introduced to improve wear in total hip arthroplasty, with manufacturers implementing different thermal treatments to reduce oxidation. It is important to understand how long-term time in vivo affects the wear of these materials. The purpose of this study was to investigate the wear and oxidative performance of first-generation HXLPE hip inserts implanted for greater than 10 years and compare annealed and remelted HXLPE formulations.

View Article and Find Full Text PDF

Aims: Micromotion of the polyethylene (PE) inlay may contribute to backside PE wear in addition to articulate wear of total knee arthroplasty (TKA). Using radiostereometric analysis (RSA) with tantalum beads in the PE inlay, we evaluated PE micromotion and its relationship to PE wear.

Methods: A total of 23 patients with a mean age of 83 years (77 to 91), were available from a RSA study on cemented TKA with Maxim tibial components (Zimmer Biomet).

View Article and Find Full Text PDF

A Polished Cobalt-Chrome Baseplate is not Associated With a Lower Revision Rate Than Matt Titanium in a Single Total Knee Arthroplasty Implant System With Identical Baseplate Design.

J Arthroplasty

April 2024

Department of Orthopaedic Surgery, St. Vincent's Hospital, The University of Melbourne, Victoria, South Australia; Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR), Adelaide, South Australia.

Background: Polished baseplates were introduced into total knee arthroplasty (TKA) systems to reduce the incidence of backside wear. In 2004, a fixed-bearing knee arthroplasty system underwent a change in baseplate material from matt titanium to polished cobalt-chrome (CoCr) with the intention to reduce backside wear. Other design aspects were left unchanged.

View Article and Find Full Text PDF

Background: Endoprosthetic distal femoral replacement (DFR) is a well-established salvage procedure following resection of malignant tumors within the distal femur. Use of an all-polyethylene tibial (APT) component is cost-effective and avoids failure due to locking-mechanism issues and backside wear, but limits modularity and the option for late liner exchange. Due to a paucity of literature we sought to answer three questions: (1) What are the most common modes of implant failure for patients undergoing cemented DFR with APT for oncologic indications? (2) What is the survivorship, rate of all-cause reoperation, and rate of revision for aseptic loosening of these implants? And (3) Is there a difference in implant survivorship or patient demographics between cemented DFRs with APT performed as a primary reconstruction those performed as a revision procedure?

Aim: To assess outcomes of cemented DFRs with APT components used for oncologic indications.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!