Background: Although guidelines for fitting hearing aids for children are well developed and have strong basis in evidence, specific protocols for fitting and verifying some technologies are not always available. One such technology is noise management in children's hearing aids. Children are frequently in high-level and/or noisy environments, and many options for noise management exist in modern hearing aids. Verification protocols are needed to define specific test signals and levels for use in clinical practice.

Purpose: This work aims to (1) describe the variation in different brands of noise reduction processors in hearing aids and the verification of these processors and (2) determine whether these differences are perceived by 13 children who have hearing loss. Finally, we aimed to develop a verification protocol for use in pediatric clinical practice.

Study Sample: A set of hearing aids was tested using both clinically available test systems and a reference system, so that the impacts of noise reduction signal processing in hearing aids could be characterized for speech in a variety of background noises. A second set of hearing aids was tested across a range of audiograms and across two clinical verification systems to characterize the variance in clinical verification measurements. Finally, a set of hearing aid recordings that varied by type of noise reduction was rated for sound quality by children with hearing loss.

Results: Significant variation across makes and models of hearing aids was observed in both the speed of noise reduction activation and the magnitude of noise reduction. Reference measures indicate that noise-only testing may overestimate noise reduction magnitude compared to speech-in-noise testing. Variation across clinical test signals was also observed, indicating that some test signals may be more successful than others for characterization of hearing aid noise reduction. Children provided different sound quality ratings across hearing aids, and for one hearing aid rated the sound quality as higher with the noise reduction system activated.

Conclusions: Implications for clinical verification systems may be that greater standardization and the use of speech-in-noise test signals may improve the quality and consistency of noise reduction verification cross clinics. A suggested clinical protocol for verification of noise management in children's hearing aids is suggested.

Download full-text PDF

Source
http://dx.doi.org/10.3766/jaaa.15060DOI Listing

Publication Analysis

Top Keywords

hearing aids
40
noise reduction
36
noise management
16
test signals
16
hearing
15
noise
12
set hearing
12
clinical verification
12
hearing aid
12
sound quality
12

Similar Publications

Objective: To compare the 3-year outcomes of the modified minimally invasive Ponto surgery (m-MIPS) to both the original MIPS (o-MIPS) and linear incision technique with soft tissue preservation (LIT-TP) for inserting bone-anchored hearing implants (BAHIs).

Study Design: Prospective study with three patient groups: m-MIPS, o-MIPS, and LIT-TP.

Setting: Tertiary referral center.

View Article and Find Full Text PDF

Detection of Extracochlear Electrodes Using Electrical Field Imaging (EFI).

Otol Neurotol

February 2025

Department of Otolaryngology-Head and Neck Surgery, Mayo Clinic, Rochester, Minnesota.

Objective: To analyze the use of electrical field imaging (EFI) in the detection of extracochlear electrodes in cochlear implants (CI).

Study Design: Retrospective cohort study.

Setting: Tertiary academic medical center.

View Article and Find Full Text PDF

Objective: The aim of this study is to test the feasibility of a custom 3D-printed guide for performing a minimally invasive cochleostomy for cochlear implantation.

Study Design: Prospective performance study.

Setting: Secondary care.

View Article and Find Full Text PDF

Objective: This study aims to evaluate the potential association of perioperative hearing outcomes with frailty by Modified 5-Item Frailty Index (mFI-5).

Design: Retrospective cross-sectional study.

Setting: Single-institutional study conducted at a tertiary care hospital between January 2018 and January 2022.

View Article and Find Full Text PDF

Objective: To investigate the association between postoperative antibiotic prophylaxis and the risk of infections leading to implant explantation or hospitalization, with a follow-up of up to 12 years.

Study Design: Retrospective cohort study.

Setting: Tertiary medical institution.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!