Purpose: Having implemented amplitude-based respiratory gating for scanned carbon-ion beam therapy, we sought to evaluate its effect on positional accuracy and throughput.
Methods And Materials: A total of 10 patients with tumors of the lung and liver participated in the first clinical trials at our center. Treatment planning was conducted with 4-dimensional computed tomography (4DCT) under free-breathing conditions. The planning target volume (PTV) was calculated by adding a 2- to 3-mm setup margin outside the clinical target volume (CTV) within the gating window. The treatment beam was on when the CTV was within the PTV. Tumor position was detected in real time with a markerless tumor tracking system using paired x-ray fluoroscopic imaging units.
Results: The patient setup error (mean ± SD) was 1.1 ± 1.2 mm/0.6 ± 0.4°. The mean internal gating accuracy (95% confidence interval [CI]) was 0.5 mm. If external gating had been applied to this treatment, the mean gating accuracy (95% CI) would have been 4.1 mm. The fluoroscopic radiation doses (mean ± SD) were 23.7 ± 21.8 mGy per beam and less than 487.5 mGy total throughout the treatment course. The setup, preparation, and irradiation times (mean ± SD) were 8.9 ± 8.2 min, 9.5 ± 4.6 min, and 4.0 ± 2.4 min, respectively. The treatment room occupation time was 36.7 ± 67.5 min.
Conclusions: Internal gating had a much higher accuracy than external gating. By the addition of a setup margin of 2 to 3 mm, internal gating positional error was less than 2.2 mm at 95% CI.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.ijrobp.2016.01.014 | DOI Listing |
Nat Commun
January 2025
Department of Electrical Engineering, University of California, Irvine, CA, USA.
Complementary transistors are critical for circuits with compatible input/output signal dynamic range and polarity. Organic electronics offer biocompatibility and conformability; however, generation of complementary organic transistors requires introduction of separate materials with inadequate stability and potential for tissue toxicity, limiting their use in biomedical applications. Here, we discovered that introduction of source/drain contact asymmetry enables spatial control of de/doping and creation of single-material complementary organic transistors from a variety of conducting polymers of both carrier types.
View Article and Find Full Text PDFEur J Orthop Surg Traumatol
January 2025
Department of Orthopaedic Surgery, University of Utah, 590 Wakara Way, Salt Lake City, UT, 84108, USA.
Purpose: Controversy remains about the ideal construct for certain olecranon fractures. The purpose of this study was to compare cost-effectiveness with the value driven outcomes tool between fixation strategies of olecranon fractures.
Methods: All surgically treated isolated proximal ulna fractures (CPT code 24,685) at a level 1 trauma center from 2013 to 2023 were retrospectively reviewed.
J Physiol
January 2025
Department of Biological Sciences, Southern Methodist University, Dallas, TX, USA.
Sudden unexpected death in epilepsy (SUDEP) is a devastating complication of epilepsy with possible sex-specific risk factors, although the exact relationship between sex and SUDEP remains unclear. To investigate this, we studied Kcna1 knockout (Kcna1) mice, which lack voltage-gated Kv1.1 channel subunits and are widely used as a SUDEP model that mirrors key features in humans.
View Article and Find Full Text PDFLancet Respir Med
December 2024
Division of Infection, Immunity and Respiratory Medicine, University of Manchester, Manchester, UK. Electronic address:
Background: The goal of cystic fibrosis transmembrane conductance regulator (CFTR) modulators is to reach normal CFTR function in people with cystic fibrosis. Vanzacaftor-tezacaftor-deutivacaftor restored CFTR function in vitro and in phase 2 trials in participants aged 18 years and older resulting in improvements in CFTR function, as measured by sweat chloride concentrations and lung function as measured by spirometry. We aimed to evaluate the efficacy and safety of vanzacaftor-tezacaftor-deutivacaftor compared with standard of care elexacaftor-tezacaftor-ivacaftor in individuals with cystic fibrosis aged 12 years and older.
View Article and Find Full Text PDFJ Imaging Inform Med
January 2025
Independent Consultant, Kirkland, WA, USA.
Point-of-care ultrasound (POCUS) has emerged as a standard of care across a variety of healthcare settings due to its ability to provide critical clinical information and as well as procedural guidance to clinicians directly at the bedside. Implementation of enterprise imaging (EI) strategies is needed such that POCUS images can be appropriately captured, indexed, managed, stored, distributed, viewed, and analyzed. Because of its unique workflow and educational requirements, reliance on traditional order-based workflow solutions may be insufficient.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!