Background: Proximal locking screw deformation and screw fracture is a frequently seen problem for femur interlocking nailing that affects fracture healing. We realized that there is lack of literature for the right level for the proximal locking screw. We investigated the difference of locking screw bending resistance between the application of screws on different proximal femoral levels.
Materials And Methods: We used a total of 80 proximal locking screws for eight groups, 10 screws for each group. Three-point bending tests were performed on four types of screws in two different trochanteric levels (the lesser trochanter and 20 mm proximal). We determined the yield points at three-point bending tests that a permanent deformation started in the locking screws using an axial compression testing machine.
Results: The mean yield point value of 5 mm threaded locking screws applied 20 mm proximal of lesser trochanter was 1022 ± 49 (range 986-1057) (mean ± standard deviation, 95% confidence interval). On the other hand, the mean yield point value of the same type of locking screws applied on the lesser trochanteric level was 2089 ± 249 (range 1911-2268). Which means 103% increase of screw resistance between two levels (P = 0.000). In all screw groups, on the lesser trochanter line we determined 98-174% higher than the yield point values of the same type of locking screws in comparison with 20 mm proximal to the lesser trochanter (P = 0.000).
Conclusion: According to our findings, there is twice as much difference in locking screw bending resistance between these two application levels. To avoid proximal locking screw deformation, locking screws should be placed in the level of the lesser trochanter in nailing of 1/3 middle and distal femur fractures.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4759882 | PMC |
http://dx.doi.org/10.4103/0019-5413.173508 | DOI Listing |
Eur J Orthop Surg Traumatol
December 2024
Department of Orthopedics and Trauma, Universidade Federal Fluminense (UFF), Niterói, Rio de Janeiro, Brazil.
Purpose: Although several techniques have been described for bent intramedullary nail removal, there is no universally accepted strategy. We hypothesized that a device based on the action principle of a three-point bend fixture could facilitate extraction of bent intramedullary nails; this paper describes its design and experimental testing.
Methods: Five large synthetic left femurs and five steel intramedullary nails were used.
BMC Musculoskelet Disord
December 2024
Department of Anatomy, Guangdong Provincial Key Laboratory of Digital Medicine and Biomechanics, Guangdong Engineering Research Center for Translation of Medical 3D Printing Application, National Virtual & Reality Experimental Education Center for Medical Morphology, School of Basic Medical Sciences, Southern Medical University, No.1023, South Shatai Road, Baiyun District, Guangzhou, Guangdong, 510515, China.
Background: This study investigated the impact of higher interfragmentary compression force (IFCF) on the stability of locking plate fixation in lateral tibial plateau fractures.
Methods: Biomechanical experiments and finite element analysis (FEA) were employed to compare the performance of the AO cancellous lag screw (AOCLS) and a newly developed combined cancellous lag screw (CCLS).
Results: The results demonstrated that the CCLS provided a higher IFCF without the risk of over-screwing, significantly improving fixation stability.
World J Surg
December 2024
Department of Cardiothoracic Surgery, The Third Hospital of Shijiazhuang City, Shijiazhuang, China.
Background: Costal cartilage injuries are unappreciated, and there is a paucity of reports on fixation methods. This study aims to evaluate the safety of titanium plate internal fixation for costal cartilage injuries.
Methods: A retrospective analysis was conducted on 30 patients with costal cartilage injuries who underwent titanium plate internal fixation between April 2016 and November 2022 at our hospital.
Background: Achieving bony union in scaphoid nonunion fractures is challenging. Various bone grafts have been studied using headless compression screws (HCS) fixation. However, the impact of bone graft choice on bone healing with volar locking plates (VLP) use is less clear.
View Article and Find Full Text PDFInjury
December 2024
Department of Orthopaedics, Larnaca General Hospital, State Health Services Organisation, Larnaca, Cyprus.
The purpose of this study was to establish typical dose values at orthopaedic operating rooms of the Larnaca General Hospital (LGH). Kerma area product (KAP), fluoroscopy time (FT) and cumulative air-kerma (K) measurements were collected for 821 patients who underwent common and reproducible trauma surgery over a five-year period, with three mobile C-arm systems; two equipped with an image-intensifier and one with a flat-panel detector. Dose indices were automatically extracted from radiation dose structured reports or DICOM meta-data files archived in the PACS, using custom-made software.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!