Purpose: This study investigated (a) the effect of amplification on cortical auditory evoked potentials (CAEPs) at different signal levels when signal-to-noise ratios (SNRs) were equated between unaided and aided conditions, and (b) the effect of absolute signal level on aided CAEPs when SNR was held constant.

Method: CAEPs were recorded from 13 young adults with normal hearing. A 1000-Hz pure tone was presented in unaided and aided conditions with a linear analog hearing aid. Direct audio input was used, allowing recorded hearing aid noise floor to be added to unaided conditions to equate SNRs between conditions. An additional stimulus was created through scaling the noise floor to study the effect of signal level.

Results: Amplification resulted in delayed N1 and P2 peak latencies relative to the unaided condition. An effect of absolute signal level (when SNR was constant) was present for aided CAEP area measures, such that larger area measures were found at higher levels.

Conclusion: Results of this study further demonstrate that factors in addition to SNR must also be considered before CAEPs can be used to clinically to measure aided thresholds.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4832873PMC
http://dx.doi.org/10.1044/2015_AJA-15-0029DOI Listing

Publication Analysis

Top Keywords

absolute signal
12
signal level
12
direct audio
8
audio input
8
unaided aided
8
aided conditions
8
hearing aid
8
noise floor
8
area measures
8
aided
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!