A New Expander for Schlemm Canal Surgery in Primary Open-angle Glaucoma-Interim Clinical Results.

J Glaucoma

*Department of Ophthalmology, Medical University of Southern Africa, Medunsa, South Africa †Department of Ophthalmology, University of Basel, Basel ‡Grieshaber Ophthalmic Research Foundation, Schaffhausen, Switzerland.

Published: August 2016

Purpose: To evaluate a new canal expander in circumferential viscocanalostomy (canaloplasty) for whites with primary open-angle glaucoma (POAG).

Design: Prospective, single-center, noncomparative study.

Participants: Twenty-two eyes of 22 consecutive patients with medically uncontrolled POAG underwent primary viscocanalostomy and implantation of the Stegmann Canal Expander into Schlemm canal with a follow-up time of at least 1 year.

Materials And Methods: Schlemm canal was unroofed ab externo, and dilated with viscoelastic material and microcatheter. The Stegmann Canal Expander is a flexible, fenestrated hollow implant of 9 mm in length. One expander was implanted into either side of the surgically created of ostium to keep the Schlemm canal permanently open. The superficial scleral flap was closed watertight. Laser goniopuncture of the trabeculo-Descemet membrane window was performed if postoperative intraocular pressure (IOP) exceeded 16 mm Hg.

Main Outcome Measures: Complete successes were defined as a confirmed IOP≤21, ≤18, and ≤16 mm Hg without medications, respectively, combined with a 30% IOP reduction. Number of antiglaucomatous medications, best-corrected visual acuity, and intraoperative and postoperative complications were recorded.

Results: The mean IOP dropped from 27.1±5.3 mm Hg preoperatively to 13.6±1.6 mm Hg at 6 months, 13.0±1.5 mm Hg at 9 months, and 13.1±2.2 mm Hg at 12 months (P<0.001). The complete success rates for an IOP≤21, ≤18, and ≤16 mm Hg combined with a 30% IOP reduction were 91% [95% confidence interval (CI), 0.80-1.0], 91% (95% CI, 0.80-1.0), and 91.0% (95% CI, 0.80-1.0) at 6 months, and 86% (95% CI, 0.73-1.0), 82% (95% CI, 0.67-1.0), and 82.0% (95% CI, 0.67-1.0) at 12 months. The success rate of an IOP≤16 mm Hg without medications did not depend on age [hazard ratios (HR) 1.02; 95% CI, 0.85-1.12; P=0.74], preoperative IOP (HR 0.97; 95% CI, 0.83-1.23; P=0.89), and mean visual defect (HR 1.05; 95% CI, 0.72-1.27; P=0.76). Laser goniopuncture was performed on 2 eyes (9%) 4.1 months postoperatively; the mean IOP was 19.5 mm Hg before and 13.6 mm Hg after goniopuncture. The number of medications dropped from 2.9±0.6 before surgery to 0.05±0.2 after surgery (P<0.001). The postoperative best-corrected visual acuity at last visit (0.12±0.09; range, 0 to 0.39) was comparable to preoperative values (mean±SD, 0.16±0.10; range, 0 to 0.39) (P=0.35). Minor intraoperative or postoperative complications included microhyphema (8 eyes) and transient elevated IOP (steroid responder; 2 eyes).

Conclusions: Implantation of the Stegmann Canal Expander in canaloplasty lowered IOP significantly in POAG without complications related to the device in this 1-year observation period.

Download full-text PDF

Source
http://dx.doi.org/10.1097/IJG.0000000000000397DOI Listing

Publication Analysis

Top Keywords

schlemm canal
16
canal expander
12
expander schlemm
8
primary open-angle
8
stegmann canal
8
canal
7
expander
5
canal surgery
4
surgery primary
4
open-angle glaucoma-interim
4

Similar Publications

Purpose: To investigate the characteristics and publication status of minimally invasive glaucoma surgery (MIGS) clinical trials registered on ClinicalTrials.gov.

Design: A cross-sectional study.

View Article and Find Full Text PDF

Purpose: Trabecular micro-bypass devices (TBDs) such as the iStent series (Glaukos Corporation, Laguna Hills, CA), are effective in reducing intraocular pressure (IOP). However, precise placement of TBDs is crucial in achieving surgical efficacy, as device malpositioning may lead to suboptimal IOP reduction. We demonstrate two novel intra-operative signs to aid confirmation of accurate iStent placement, without reliance on imaging technologies.

View Article and Find Full Text PDF

Introduction: Cataract surgery has been reported to have a reducing effect on intraocular pressure (IOP) in glaucomatous and non-glaucomatous eyes. This effect seems to be more noticeable in eyes with narrow angles (NAs) than in eyes with open angles (OAs). Decrease in IOP may be a result of the increase in anterior chamber angle (ACA) and Schlemm canal (SC) after cataract surgery.

View Article and Find Full Text PDF

Since the trabecular meshwork (TM) is central to intraocular pressure (IOP) regulation and glaucoma, a deeper understanding of its genomic landscape is needed. We present a multimodal, single-cell resolution analysis of mouse limbal cells (includes TM). In total, we sequenced 9,394 wild-type TM cell transcriptomes.

View Article and Find Full Text PDF

Minimally invasive glaucoma surgeries (MIGS) offer an alternative approach, aiming to be both safe and effective. Defined by their ab interno technique, which avoids creating a bleb, MIGS procedures seek to reduce intraocular pressure (IOP) by enhancing the outflow pathways of aqueous humor. Techniques include stenting, incisions, and excisions of the trabecular meshwork, canaloplasty of Schlemm's canal, and drainage implants that enhance uveoscleral outflow to the subchoroidal space.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!