Study Design: Cross-sectional study.

Purpose: To accurately measure the dimensions of cervical endplates based on computed tomography (CT) scans in Indian population and assess accuracy of match with currently available cervical disc prostheses.

Overview Of Literature: The dimensions of currently available cervical disc replacement implants are based on early published geometrical measurements of vertebrae endplates for Caucasian population. To author's knowledge, similar study has not been published for patients from Indian subcontinent.

Methods: CT scans of cervical spine of patients from Indian subcontinent were collected and reviewed. Seventy patients (54 men and 16 women; aged 18-56 years with average of 37 years) who underwent CT scans of cervical spine were included in study. 3D CT scans of sub axial cervical spine (C3 to C7) were analyzed. The anterior-posterior (AP) and central mediolateral (CML) dimensions of superior and inferior endplates from C3 to C7 were measured using digital measuring system.

Results: A total of 560 endplates of 70 patients were included in the study. The AP diameter of cervical endplates ranged from 0.87 to 2.47 cm. The CML diameters ranged from 0.84 to 2.98 cm. For levels C3/C4 and C4/C5 for AP dimension Prestige-LP (90.5%) and Prodisc-C (89%) discs showed higher percentage of matching than Discover discs (58.5%). For CML diameter, Prestige-LP (69.5%), Prodisc-C (70%) and Discover (39.5%) discs showed almost similar matching with measured endplates. For levels C5/C6 and C6/C7 for AP dimension, Prestige-LP (67.25%), Prodisc-C (49.35%) and Discover (51.5%) discs showed similar matching. For CML diameter Prestige-LP (32%), Prodisc-C (27.5%) and Discover (42.2%) discs showed poor matching with measured endplates.

Conclusions: This study indicates need for redesign of cervical disc prostheses to match Indian patients. The collected anthropometric dimensions from this study may be used to design and develop indigenous artificial total disc replacement prosthesis and even cervical cages in India. With the present study being a small pilot study, the authors recommend anthropometric CT measurements in larger number of Indian patients in order to validate footprint dimensions for designing better-matched prosthesis.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4764536PMC
http://dx.doi.org/10.4184/asj.2016.10.1.20DOI Listing

Publication Analysis

Top Keywords

disc replacement
12
cervical disc
12
cervical spine
12
cervical
10
indian population
8
study
8
study design
8
cervical endplates
8
currently cervical
8
patients indian
8

Similar Publications

Background: Secondary mitral regurgitation (SMR) is a condition affecting the left ventricle (LV) rather than the mitral valve (MV). If the MV remains structurally unchanged, enlargement of the LV or impairment of the papillary muscles can occur. Several mechanical interventions are available to dictate the resolution of MR.

View Article and Find Full Text PDF

Secondary mitral regurgitation (SMR) is characterized by a pathological process impacting the left ventricle (LV) as opposed to the mitral valve (MV). In the absence of structural alterations to the MV, the expansion of the LV or impairment of the papillary muscles (PMs) may ensue. A number of technical procedures are accessible for the purpose of determining the optimal resolution for MR.

View Article and Find Full Text PDF

Study Design: Matched case-control study.

Purpose: To evaluate the midterm outcomes of unilateral pedicle screw fixation (UPSF) versus bilateral pedicle screw fixation (BPSF) in transforaminal lumbar interbody fusion (TLIF) procedure, ascertain efficacy of UPSF in adequately decompressing contralateral foramen+spinal canal and reducing rate of adjacent segment degeneration (ASD) at 4-8-year follow-up (FU).

Overview Of Literature: Previous meta-analyses found no significant differences between UPSF and BPSF regarding fusion rates, clinical and radiological outcomes; however, few studies have reported higher rates of cage migration/subsidence and pseudoarthrosis in the UPSF.

View Article and Find Full Text PDF

Study Design: A retrospective comparative study.

Purpose: To validate the hypothesis that a combination of multilevel Ponte osteotomy (PO) with intraoperative traction (IOT) results in a better correction than IOT alone in high-magnitude curves in adolescent idiopathic scoliosis (AIS) and does not possess an attributable risk of neurological injury.

Overview Of Literature: On a comprehensive review of the literature, the choice of technique adopted for curves between 65° and 100° remains controversial with no major consensus favoring one technique over the other.

View Article and Find Full Text PDF

Device profile of the Mobi-C artificial cervical disc: an overview of its safety and efficacy.

Expert Rev Med Devices

January 2025

Department of Orthopaedic Surgery, UC Davis Health, Sacramento, California, USA.

Article Synopsis
  • The Mobi-C Cervical Disc Replacement is FDA-approved for treating multiple levels of the cervical spine and is shown to be effective compared to traditional fusion surgery based on clinical trial data.
  • The article reviews the history, biomechanics, clinical results, long-term outcomes, and cost-effectiveness of Mobi-C, indicating it is a strong alternative to fusion.
  • Expert opinion highlights Mobi-C's proven track record, suggesting it preserves motion, provides positive clinical outcomes, and ensures patient safety, making it a favorable option for appropriate candidates.
View Article and Find Full Text PDF