Purpose: To characterize the performance of a new unshielded silicon diode (Razor-IBA) for dose measurements in small flattening filter free beams.
Methods: The Razor has an active volume of 0.6 mm in diameter and 20 µm in length. The detector response stability in measured dose, dose rate, dose per pulse, and dark current were evaluated. The detector response in square fields (0.6-5.0 cm) was determined using PDD curves, axial beam profiles and output ratios. The performances were compared to that of the previously available SFD-IBA and PFD-IBA diodes.
Results And Discussion: The Razor short term stability relative to the SFD was much improved (<±0.1% after 1.2 kGy). The linearity was <±1% (0.05-30 Gy range) and the dose rate dependence was <±0.5% (4-24 Gy/min range). The dose per pulse dependence was <±0.7% (0.08-0.21 cGy/pulse range). The PDDs measured with Razor and PFD differed <1%. A larger dark current was observed with increase in dose (0.0025 pA/Gy) compared to the SFD (0.0002 pA/Gy). This characteristic is attributed to an increased concentration of recombination centers. The beam profile showed good agreement with the SFD. Penumbra differences were <±0.3 mm relative to PFD, with a slight overestimation of the tails (<1%), due to the absence of diode shielding. Output ratios were in good agreement for fields up to 5 × 5 cm(2) (<1%).
Conclusions: The Razor diode has the same spatial resolution and performance reliability as its predecessor (SFD), but exhibits the additional advantage of improved stability. These features make the Razor diode detector a good candidate for small field dosimetry.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.ejmp.2016.02.004 | DOI Listing |
Proc Natl Acad Sci U S A
January 2025
Department of Ecology and Evolutionary Biology, University of Kansas, Lawrence, KS 66045.
Climate change is increasing the frequency of large-scale, extreme environmental events and flattening environmental gradients. Whether such changes will cause spatially synchronous, large-scale population declines depends on mechanisms that limit metapopulation synchrony, thereby promoting rescue effects and stability. Using long-term data and empirical dynamic models, we quantified spatial heterogeneity in density dependence, spatial heterogeneity in environmental responses, and environmental gradients to assess their role in inhibiting synchrony across 36 marine fish and invertebrate species.
View Article and Find Full Text PDFJ Appl Clin Med Phys
January 2025
Department of Radiotherapy, University Medical Center Utrecht, Utrecht, Netherlands.
Introduction: This paper describes a method to improve gantry-dependent beam steering for Elekta traveling wave linear accelerators by applying the measured and filtered beam servo corrections to the existing lookup table (LUT). Beam steering has a direct influence on the treatment accuracy by affecting the beam symmetry and position. The presented method provides an improved LUT with respect to the default Elekta method to reduce treatment delivery interruptions.
View Article and Find Full Text PDFPituitary
December 2024
Dipartimento di Medicina Traslazionale, Università Cattolica del Sacro Cuore, Rome, Italy.
Introduction: Empty sella is characterized by a flattened profile of the pituitary gland that represents in most cases only a radiological incidental finding. When endocrine, ophthalmic, and neurological symptoms occur, this condition is described as empty sella syndrome.
Materials And Methods: We searched MEDLINE (PubMed database) with the data filter 2024-2009 using the keywords listed above.
Med Phys
December 2024
Department of Physics, The Verspeeten Family Cancer Centre, London, Ontario, Canada.
Background: Stereotactic arrythmia radioablation (STAR) is a noninvasive technique to treat ventricular tachycardia (VT). Management of cardiorespiratory motion plays an essential role in VT-STAR treatments to improve treatment outcomes by reducing positional uncertainties and increasing dose conformality. Use of an electrocardiogram (ECG) signal, acquired in real-time, as a surrogate to gate the beam has the potential to fulfil that intent.
View Article and Find Full Text PDFMed Phys
December 2024
Division of Ionizing Radiation Metrology, National Institute of Metrology (NIM), Beijing, China.
Background: The clinical use of flattening filter free (FFF) radiotherapy has significantly increased in recent years due to its effective enhancement of dose rates and reduction of scatter dose. A proposal has been made to adjust the incident electron angle of the accelerator to expand the application of FFF beams in areas such as large planning target volumes (PTVs). However, the inherent softening characteristics and non-uniformity of lateral dose distribution in FFF beams inevitably lead to increased dosimetry errors, especially for ionization chambers widely used in clinical practice, which may result in serious accidents during FFF radiotherapy.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!