Background: Femorotibial kinematics and contact patterns vary greatly with different total knee arthroplasty (TKA) designs. Therefore, guided motion knee systems were developed to restore natural knee kinematics and make them more predictable. The medial stabilized TKA design is supposed to replicate physiological kinematics more than the posterior-stabilized TKA system. We conducted this study to compare a newly developed medial stabilized design with a conventional posterior-stabilized design in terms of femorotibial kinematics and contact patterns in vitro.
Methods: Twelve fresh-frozen knee specimens were tested in a weight-bearing knee rig after implantation of a posterior stabilized and medial-stabilized total knee arthroplasty under a loaded squat from 20° to 120° of flexion. Femorotibial joint contact pressures in the medial and lateral compartments were measured by pressure sensitive films and knee kinematics were recorded by an ultrasonic 3-dimensional motion analysis system.
Findings: The medial stabilized design showed a reduction of medial femorotibial translation compared to posterior-stabilized design (mean 3.5mm compared to 15.7 mm, P<0.01). In the lateral compartment, both designs showed a posterior translation of the femur with flexion, but less in the medial stabilized design (mean 14.7 mm compared to 19.0mm, P<0.01). In the medial femorotibial compartment of medial stabilized design, we observed an enlarged contact area and lower peak pressure, in contrast in the lateral compartment there was a reduced contact area and an increased peak pressure.
Interpretation: While posterior-stabilized design enforces a medio-lateral posterior translation, the medial stabilized arthroplasty system enables a combination of a lateral translation with a medial pivot, which restores the physiological knee kinematics better.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.clinbiomech.2016.02.002 | DOI Listing |
J Exp Orthop
January 2025
Department of Orthopaedic Surgery, Faculty of Medicine The University of Tokyo Tokyo Japan.
Purpose: To clarify the influence of biomechanics on post-operative clinical outcomes in bicruciate-retaining total knee arthroplasty (BCR-TKA).
Methods: Severe medial osteoarthritis who underwent BCR-TKA were examined. Each patient was asked to perform a squat (weight-bearing [WB]) and active assisted knee flexion (non-WB [NWB]) under single fluoroscopy surveillance.
JBJS Case Connect
January 2025
Morinomiya University of Medical Sciences, Graduate School of Health Sciences, Osaka, Japan.
Case: A 75-year-old woman underwent cruciate retaining total knee arthroplasty (TKA) for osteonecrosis of the knee, converted from unicompartmental arthroplasty due to an intraoperative medial tibial plateau fracture. Four years later, she experienced persistent knee pain. Computed tomography revealed tibial malalignments in 2 planes.
View Article and Find Full Text PDFJ Arthroplasty
October 2024
Center for Musculoskeletal Research, University of Tennessee-Knoxville, Knoxville, Tennessee.
Background: The non-implanted knee differs in comparison to total knee arthroplasty (TKA) designs, with regard to asymmetry and functionality of the anterior cruciate ligament and the posterior cruciate ligament. While surgeons may choose to implant either posterior stabilized (PS) or bi-cruciate stabilized (BCS) TKAs, substituting for one or both cruciate ligaments, the effects of symmetry versus asymmetry in substituting TKA designs have not been widely analyzed to determine possible benefits. Therefore, the objective of this research study was to determine if either TKA asymmetry and/or anterior ligament stabilization can lead to more normal-like kinematics and clinical benefit for patients.
View Article and Find Full Text PDFBioengineering (Basel)
September 2024
Department of Biomedical Engineering, Department of Mechanical Engineering, Department of Orthopaedic Surgery, University of California at Davis, One Shields Avenue, Davis, CA 95616, USA.
Current surgical practices in total knee arthroplasty (TKA) have advanced and include significant changes and improvements in alignment philosophies, femorotibial implant conformities, and ligament management to replicate in vivo knee kinematics. While corrective measures have emphasized sagittal plane alignment to restore normal flexion-extension (F-E) motion and coronal plane ligament balance, internal-external (I-E) rotation kinematics in the axial plane have been largely neglected. Recent in vivo evidence indicates that the combination of factors necessary to closely restore native tibial rotation as the knee flexes and extends is kinematic alignment (KA), which resurfaces the patient's pre-arthritic knee without releasing ligaments, an insert with medial 1:1 ball-in-socket conformity and a lateral flat surface, and posterior cruciate ligament (PCL) retention.
View Article and Find Full Text PDFScand J Med Sci Sports
September 2024
Department of Natural Sciences in Kinanthropology, Faculty of Physical Culture, Palacky University Olomouc, Olomouc, Czech Republic.
The aim of the present study was to examine the effects of attentional focus instructions on acute changes in the transverse relaxation time (T2) of the femorotibial cartilage and in cartilage volume during repeated drop-jump landings. Ten healthy females (M = 20.4 ± 0.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!