Purpose: To assess the accuracy of best-corrected visual acuity (BCVA) measured by non-ophthalmic emergency department (ED) staff with a standard Snellen chart versus an automated application (app) on a handheld smartphone (Paxos Checkup, San Francisco, CA, USA).
Methods: The study included 128 subjects who presented to the Stanford Hospital ED for whom the ED requested an ophthalmology consultation. We conducted the study in two phases. During phase 1 of the study, ED staff tested patient BCVA using a standard Snellen test at 20 feet. During phase 2 of the study, ED staff tested patient near BCVA using the app. During both phases, ophthalmologists measured BCVA with a Rosenbaum near chart, which was treated as the gold standard. ED BCVA measurements were benchmarked prospectively against ophthalmologists' measurements and converted to logMAR.
Results: ED logMAR BCVA was 0.21 ± 0.35 (approximately 2 Snellen lines difference ± 3 Snellen lines) higher than that of ophthalmologists when ED staff used a Snellen chart (p = .0.00003). ED BCVA was 0.06 ± 0.40 (less than 1 Snellen line ± 4 Snellen lines) higher when ED staff used the app (p = 0.246). Inter-observer difference was therefore smaller by more than 1 line (0.15 logMAR) with the app (p = 0.046).
Conclusions: BCVA measured by non-ophthalmic ED staff with an app was more accurate than with a Snellen chart. Automated apps may provide a means to standardize and improve the efficiency of ED ophthalmologic care.