A Laser-Assisted Anastomotic Technique: Feasibility on Human Diseased Coronary Arteries.

Innovations (Phila)

From the *Department of Cardiothoracic Surgery, University Medical Center Utrecht, Utrecht, the Netherlands; †Corvasco Medical B.V., Utrecht, the Netherlands; and ‡Department of Pathology, Laboratories and Pharmacy Division, §Department of Neurosurgery, Brain Division, and ||Department of Experimental Cardiology, Heart and Lungs Division, University Medical Center Utrecht, Utrecht, the Netherlands.

Published: May 2017

AI Article Synopsis

  • The study evaluates the efficacy of Excimer laser-assisted Trinity Clip anastomotic connectors in human diseased coronary arteries and assesses potential acute laser effects on the coronary wall.
  • Study 1 involved 38 ex vivo anastomoses where connector positioning was affected by the severity of atherosclerosis, while Study 2 examined 20 anastomoses for laser-induced damage in different models.
  • Results showed high success rates for laser-punching on early lesions, but complications arose with severely diseased walls; however, no histological damage was seen when used correctly, suggesting careful coronary selection is crucial for success.

Article Abstract

Objective: Atherosclerotic disease might hamper the efficacy of the Excimer laser-assisted Trinity Clip anastomotic connector in coronary arteries. Therefore, its efficacy was evaluated on human diseased coronary arteries (study 1). In addition, the acute laser effects onto the coronary wall were assessed (study 2).

Methods: Thirty-eight anastomoses were constructed on ex vivo human hearts. Atherosclerosis was histopathologically determined and subsequently related to the success of the technique (ie, connector positioning and laser punching; study 1). In addition, 20 anastomoses were constructed in an ex vivo (porcine, n = 8) and an in vivo [rabbit (n = 9) and porcine (n = 3)] model. Subsequently, the coronary was histologically studied on the presence of laser-induced damage (study 2).

Results: In 13 of 38 anastomoses (study 1), the connector was malpositioned, 3 because of a severely diseased coronary wall and 10 because of an inner diameter less than the intended target range. The laser-punch success rates on coronary arteries with an early and advanced lesion were 100% (16/16) and 89% (8/9; lesions were located in the inferolateral wall), respectively. In one case, an advanced lesion (ie, fibrocalcified plaque) was located in the superolateral wall and caused a laser-punch failure. No histological signs of laser-induced damage were observed, in case of correct use (study 2).

Conclusions: This study demonstrates the feasibility of an anastomotic connector on human diseased coronary arteries and shows that lasering does not induce coronary wall damage. However, careful selection of the coronary, regarding the target inner diameter and disease status, will prevent construction failures. This connector could facilitate less invasive coronary artery bypass grafting.

Download full-text PDF

Source
http://dx.doi.org/10.1097/IMI.0000000000000241DOI Listing

Publication Analysis

Top Keywords

coronary arteries
20
diseased coronary
16
human diseased
12
coronary wall
12
coronary
11
anastomotic connector
8
study addition
8
anastomoses constructed
8
constructed vivo
8
laser-induced damage
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!