Objective: To compare the biomechanical properties of locking compression plate (LCP) and a limited contact dynamic compression plate combined with an intramedullary rod (LC-DCP-R) in a cadaveric, canine, femoral fracture-gap model.

Study Design: In vitro biomechanical study; nonrandomized, complete block (dog).

Sample Population: Paired cadaveric canine femora (n = 10 dogs).

Methods: Paired femurs with a mid-diaphyseal 20 mm gap were stabilized with either LCP or LC-DCP-R. Nondestructive testing up to 60% of body weight (BW) was followed by a continuous destructive test. Comparative structural properties, 3-dimensional (3D) interfragmentary motion, and plate linear strain were evaluated. Paired comparisons were made between LCP and LC-DCP-R.

Results: Stiffness after nondestructive testing was significantly lower for LCP with a mean (95% confidence interval [CI]) of 61 N/mm (46-76) versus 89 N/mm (67-110) for LC-DCP-R (P = .0072). Ultimate load to failure was significantly lower for LCP with a median (interquartile range [IQR]) of 270 N (247-286) versus 371.5 (353-385) for LC-DCP-R (P = .002). Axial motion at 60% BW was significantly higher for LCP with a median (IQR) of 1.01 mm (0.71-1.26) versus 0.36 mm (0.20-0.49) for LC-DCP-R (P = .002). Shear motion was significantly higher for LCP with a median (IQR) of 1.18 (0.78-1.58) versus 0.72 mm (0.45-1.00) for LC-DCP-R (P = .018). Strain was significantly higher for mid-LCP surface with a mean (95%CI) at 60% BW of 979 μdef (579-1378) versus 583 μdef (365-801) at mid-LC-DCP-R surface (P = .0153). The elastic limit strain of the plates was not different and was reached at a mean (95%CI) load of 241 N (190-292) for LCP versus 290 N (245-336) for LC-DCP-R (P = .12).

Conclusion: The LC-DCP-R showed higher stiffness and resistance to failure, lower interfragmentary motion, and lower plate strain and stress compared to LCP.

Download full-text PDF

Source
http://dx.doi.org/10.1111/vsu.12451DOI Listing

Publication Analysis

Top Keywords

compression plate
16
lcp median
12
lcp
9
locking compression
8
limited contact
8
contact dynamic
8
dynamic compression
8
plate combined
8
combined intramedullary
8
intramedullary rod
8

Similar Publications

Objective: To observe the clinical outcomes of anterior approach for the revision surgery following unsuccessful bone cement augmentation in osteoporotic vertebral compression fractures.

Methods: A total of 10 patients who experienced unsuccessful bone cement augmentation underwent anterior revision surgery between January 2020 and December 2021. There were 2 males and 8 females.

View Article and Find Full Text PDF

Preparation of Hydroxyapatite-Aligned Collagen Sheets and Their Evaluation for Fibroblast Adhesion and Collagen Secretion.

ACS Biomater Sci Eng

January 2025

Department of Materials Science and Engineering, School of Materials and Chemical Technology, Institute of Science Tokyo, 2-12-1 Ookayama, Meguro-ku, Tokyo 152-8550, Japan.

The structure of many native tissues consists of aligned collagen (Col) fibrils, some of which are further composited with dispersed hydroxyapatite (HAp) nanocrystals. Accurately mimicking this inherent structure is a promising approach to enhance scaffold biocompatibility in tissue engineering. In this study, biomimetic sheets composed of highly aligned Col fibrils were fabricated using a plastic compression and tension method, followed by the deposition of HAp nanocrystals on the surface via an alternate soaking method.

View Article and Find Full Text PDF

We demonstrate a high-performance ultrafast broadband time-resolved photoluminescence (TRPL) system based on the transient grating photoluminescence spectroscopy (TGPLS) technique. The core of the system is a Kerr effect-induced transient grating (TG) optical gate driven by high repetition rate ultrashort laser pulses at 1030 nm with micro-Joule pulse energy. Satisfying the demands of spectroscopy applications, the setup achieves high sensitivity, rapid data acquisition, ultrafast time resolution, and a wide spectral window from ultraviolet to near-infrared.

View Article and Find Full Text PDF

Locking plates have a rapidly growing process especially in the past decades and results are satisfactory especially in the osteoporotic bones compared to non-locking compression plates. There are many forms of failure in the fracture fixation of locking plates, and screw pull-out is one of the main failure reasons. In this study, we aim to investigate pull-out failure in locking plates using locking spongious screws.

View Article and Find Full Text PDF

Objective:  To determine the effect of locking head inserts (LHI) on plate strain, stiffness, and deformation when applied to a 3.5-mm broad locking compression plate (LCP) in an open fracture-gap model.

Study Design:  Six, 13-hole, 3.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!