New mitral annular force transducer optimized to distinguish annular segments and multi-plane forces.

J Biomech

Department of Cardiothoracic and Vascular Surgery and Department of Clinical Medicine, Aarhus University Hospital, Aarhus, Denmark; Department of Biomedical Engineering, University of Arkansas, Fayetteville, AR, USA. Electronic address:

Published: March 2016

Limited knowledge exists about the forces acting on mitral valve annuloplasty repair devices. The aim of this study was to develop a new mitral annular force transducer to measure the forces acting on clinically used mitral valve annuloplasty devices. The design of an X-shaped transducer in the present study was optimized for simultaneous in- and out-of-plane force measurements. Each arm was mounted with strain gauges on four circumferential elements to measure out-of-plane forces, and the central parts of the X-arms were mounted with two strain gauges to measure in-plane forces. A dedicated calibration setup was developed to calibrate isolated forces with tension and compression for in- and out-of-plane measurements. With this setup, it was possible with linear equations to isolate and distinguish measured forces between the two planes and minimize transducer arm crosstalk. An in-vitro test was performed to verify the crosstalk elimination method and the assumptions behind it. The force transducer was implanted and evaluated in an 80kg porcine in-vivo model. Following crosstalk elimination, in-plane systolic force accumulation was found to be in average 4.0±0.1N and the out-of-plane annular segments experienced an average force of 1.4±0.4N. Directions of the systolic out-of-plane forces indicated movements towards a saddle shaped annulus, and the transducer was able to measure independent directional forces in individual annular segments. Further measurements with the new transducer coupled with clinical annuloplasty rings will provide a detailed insight into the biomechanical dynamics of these devices.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jbiomech.2016.02.003DOI Listing

Publication Analysis

Top Keywords

force transducer
12
annular segments
12
forces
9
mitral annular
8
annular force
8
forces acting
8
mitral valve
8
valve annuloplasty
8
transducer measure
8
in- out-of-plane
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!