A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

A New 35-mm Short Intra-aortic Balloon Catheter: A Suitable Option Also for Non-Small-Sized Patients? | LitMetric

A New 35-mm Short Intra-aortic Balloon Catheter: A Suitable Option Also for Non-Small-Sized Patients?

Innovations (Phila)

From the *Maastricht University Medical Center, Maastricht, the Netherlands; †Careggi Hospital, Florence, Italy; and ‡Department of Cardiothoracic Surgery, University Hospital, Brussels, Belgium.

Published: January 2017

Objective: Visceral ischemia can be a potentially life-threatening complication of intra-aortic balloon pump (IABP) support. A shorter IABP catheter might lead to a reduction of visceral complications. In this animal study, we evaluate the effects of a 35-mL short catheter in comparison with a 40-mL standard-sized catheter.

Methods: Eighteen healthy swine underwent 120-minute ligation of the left anterior descending coronary artery followed by 6 hours of reperfusion being supported by either a short IABP catheter (short group) (n = 6) or a long IABP catheter (long group) (n = 6) or with no assistance (controls) (n = 6). Hemodynamics, visceral and coronary flows, as well as biochemical markers were evaluated throughout the different phases of the protocol.

Results: Mesenteric flows increased significantly at reperfusion (P < 0.001 both) remaining constant afterward (all, P > 0.05) in the short group, while remaining significantly lower in the long group at the start of reperfusion, remaining constantly lower than the short group and controls (P < 0.001 vs short, P < 0.003 vs controls). In both long and short groups, catheters improved renal flows at reperfusion (P < 0.001 both) without any further variation (P > 0.05). In the short group, the flows were higher during the whole of reperfusion (all, P < 0.05). Intra-aortic balloon pump support improved hemodynamic indices and coronary blood flows during reperfusion to a similar extent in both the small and the long group (P > 0.05).

Conclusions: The short IABP catheter proved to be as effective as the standard-sized catheter in supporting hemodynamics and coronary circulation. Furthermore, it even improves visceral flows in comparison with conventional IABP catheters.

Download full-text PDF

Source
http://dx.doi.org/10.1097/IMI.0000000000000233DOI Listing

Publication Analysis

Top Keywords

iabp catheter
16
short group
16
intra-aortic balloon
12
long group
12
short
9
balloon pump
8
short iabp
8
reperfusion 0001
8
005 short
8
flows reperfusion
8