A comparative MS study was conducted on the analytical performance of two matrix-assisted laser desorption/ionization (MALDI) sources that operated at either low pressure (∼1Torr) or at atmospheric pressure. In both cases, the MALDI sources were attached to a linear ion trap mass spectrometer equipped with a two-stage ion funnel. The obtained results indicate that the limits of detection, in the analysis of identical peptide samples, were much lower with the source that was operated slightly below the 1-Torr pressure. In the low-pressure (LP) MALDI source, ion signals were observed at a laser fluence that was considerably lower than the one determining the appearance of ion signals in the atmospheric pressure (AP) MALDI source. When the near-threshold laser fluences were used to record MALDI MS spectra at 1-Torr and 750-Torr pressures, the level of chemical noise at the 1-Torr pressure was much lower compared to that at AP. The dependency of the analyte ion signals on the accelerating field which dragged the ions from the MALDI plate to the MS analyzer are presented for the LP and AP MALDI sources. The study indicates that the laser fluence, background gas pressure, and field accelerating the ions away from a MALDI plate were the main parameters which determined the ion yield, signal-to-noise (S/N) ratios, the fragmentation of the analyte ions, and adduct formation in the LP and AP MALDI MS methods. The presented results can be helpful for a deeper insight into the mechanisms responsible for the ion formation in MALDI.