A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Decreased Hemolysis and Improved Hemodynamic Performance of Synchronized Bileaflet Mechanical Valve. | LitMetric

Decreased Hemolysis and Improved Hemodynamic Performance of Synchronized Bileaflet Mechanical Valve.

Ann Thorac Surg

Division of Cardiology, Department of Internal Medicine, Chang Gung Memorial Hospital at Keelung, Keelung, Taiwan, Republic of China; College of Medicine, Chang Gung University, Tao-Yuan, Taiwan, Republic of China.

Published: March 2016

Purpose: This study compared the performance of a newly patented synchronized valve (SV) with that of a commercially available (CAV) bileaflet mechanical heart valve.

Description: A high-speed camera was used to record the leaflet kinematics of the SV vs the CAV along the flow channel. Transvalvular energy loss, effective orifice area, and hemolysis ratios were obtained using a mock circulatory system at two fixed pulse rates and at various cardiac outputs with a fixed aortic pressure.

Evaluation: The rotational radius and inertia of the SV was lower than that of the CAV during valve closure. For heart rates and at cardiac outputs of 7, 5, and 4 L/min, the ratio of total energy loss to effective energy of the SV was significantly less than the CAV, whereas the effective orifice area of the SV was significantly larger than that of CAV. The hemolysis ratio after 4 hours was significantly higher in the CAV than in the SV for both pulse rates.

Conclusions: The synchronized leaflet motion mitigated leaflet rebound and regurgitation during valve closure, which could decrease energy loss, increase the effective orifice area, and reduce hemolysis.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.athoracsur.2015.10.111DOI Listing

Publication Analysis

Top Keywords

energy loss
12
effective orifice
12
orifice area
12
bileaflet mechanical
8
loss effective
8
rates cardiac
8
cardiac outputs
8
valve closure
8
cav
6
decreased hemolysis
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!