The Elekta Versa HD incorporates a variety of upgrades to the line of Elekta linear accelerators, primarily including the Agility head and flattening filter-free (FFF) photon beam delivery. The completely distinct dosimetric output of the head from its predecessors, combined with the FFF beams, requires a new investigation of modeling in treatment planning systems. A model was created in Pinnacle3 v9.8 with the commissioned beam data. A phantom consisting of several plastic water and Styrofoam slabs was scanned and imported into Pinnacle3, where beams of different field sizes, source-to-surface distances (SSDs), wedges, and gantry angles were devised. Beams included all of the available photon energies (6, 10, 18, 6FFF, and 10 FFF MV), as well as the four electron energies commissioned for clinical use (6, 9, 12, and 15 MeV). The plans were verified at calculation points by measurement with a calibrated ionization chamber. Homogeneous and hetero-geneous point-dose measurements agreed within 2% relative to maximum dose for all photon and electron beams. AP photon open field measurements along the central axis at 100 cm SSD passed within 1%. In addition, IMRT testing was also performed with three standard plans (step and shoot IMRT, as well as a small- and large-field VMAT plan). The IMRT plans were delivered on the Delta4 IMRT QA phantom, for which a gamma passing rate was > 99.5% for all plans with a 3% dose deviation, 3 mm distance-to-agreement, and 10% dose threshold. The IMRT QA results for the first 23 patients yielded gamma passing rates of 97.4% ± 2.3%. Such testing ensures confidence in the ability of Pinnacle3 to model photon and electron beams with the Agility head.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5690210PMC
http://dx.doi.org/10.1120/jacmp.v17i1.5808DOI Listing

Publication Analysis

Top Keywords

agility head
12
head flattening
8
flattening filter-free
8
photon electron
8
electron beams
8
gamma passing
8
photon
5
beams
5
imrt
5
pinnacle3
4

Similar Publications

Background/objectives: Physical fitness and diet along with body weight are key determinants of health. Excess body weight, poor dietary choices, and low physical fitness, however, are becoming increasingly prevalent in adolescents. In order to develop adequate intervention strategies, additional research on potential interaction effects of these entities is needed.

View Article and Find Full Text PDF

Background: Physical fitness (PF) is important for children's physical and cognitive development. There is increasing interest in the relationship between physical fitness, and executive function. Since it is unclear which physical fitness component benefits which domain of executive functioning, it's challenging to develop effective physical intervention programs for children with executive functioning disorders.

View Article and Find Full Text PDF

Introduction Volumetric-modulated arcs (VMA) can produce dose distributions suitable for stereotactic radiosurgery (SRS) with a multi-leaf collimator (MLC) for brain metastases (BMs). The treatment planning and verification for VMA are more complicated than for dynamic conformal arcs. The longer the preparation time from image acquisition to the start of irradiation, the higher the risk of tumor growth and/or displacement.

View Article and Find Full Text PDF
Article Synopsis
  • The ATP-PCr system is crucial for high-intensity actions in taekwondo, while the glycolytic system helps sustain these efforts during repeated techniques.
  • This narrative review focuses on sport-specific anaerobic tests for taekwondo, analyzing 41 studies published from 2014 to 2023 to highlight various testing protocols.
  • Tests include the Taekwondo Anaerobic Test and various kick tests, designed for easy implementation by coaches and trainers to monitor and improve anaerobic performance, with guidelines on their validity, reliability, and application.
View Article and Find Full Text PDF
Article Synopsis
  • This study compares Monte Carlo (MC) and Collapsed Cone (CC) dose algorithms in RayStation 12A for 6 MV and 6 MV flattening filter-free (FFF) photon beams used in radiotherapy plans involving small and highly modulated fields.
  • It assesses how these algorithms perform in output dose measurements, examining their accuracy across various configurations such as rectangular and complex clinical plans.
  • Results indicate that while both algorithms show dependency on collimator angles in narrow fields, MC may perform better in highly modulated scenarios, but is less reliable with inhomogeneous ArcCHECK images, suggesting the need for homogeneous phantom setups for verification.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!