Purpose: To examine the changes in distraction force following relining of a conventional abdominal aortic stent-graft with a type IIIb endoleak using the Nellix endovascular sealing device compared to a unilateral stent-graft.

Methods: Relining is often used to repair type IIIb endoleaks, but the consequences to graft stability are unknown. A mathematical model was constructed based on pressure and volume flow through the stent-grafts, incorporating recognized distraction force equations. Steady flow was presumed at peak systolic pressures to calculate the maximum distraction force, with gravity ignored. Distraction forces for 28- to 36-mm-diameter stent-graft bodies with 16-mm limbs were calculated and compared to forces following relining with single and double Nellix devices or the Renu unilateral device.

Results: Distraction forces for the 28-, 32-, and 36-mm stent-grafts prior to relining were 5.99, 10.21, and 14.99 N, respectively. Similar forces were reported after relining with bilateral Nellix devices (5.86, 10.08, and 14.86 N, respectively). However, use of a unilateral Nellix increased the distraction forces to 9.92, 14.14, and 18.92 N, respectively. These were comparable to the increase observed after relining with a Renu unilateral stent-graft (9.87, 14.09, and 18.86 N, respectively). The proportional increase in distraction force for a unilateral relining ranged from 26% to 66%, with the greatest increase noted in the smaller diameter main bodies.

Conclusion: Relining a stent-graft with a type IIIb endoleak using bilateral Nellix devices does not increase the distraction force. However, a unilateral Nellix device or the Renu system could theoretically increase the distraction force by up to 66%, potentially risking migration and type Ia endoleak. In clinical practice, these results suggest that a relining with bilateral Nellix may have benefits over the Renu unilateral stent-graft.

Download full-text PDF

Source
http://dx.doi.org/10.1177/1526602816631106DOI Listing

Publication Analysis

Top Keywords

distraction force
24
type iiib
16
distraction forces
16
iiib endoleak
12
nellix devices
12
renu unilateral
12
bilateral nellix
12
increase distraction
12
relining
10
distraction
10

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!