Background: In this study, the accuracy of ultra-low-dose computed tomography (CT) with iterative reconstruction (IR) for detection and measurement of pulmonary nodules was evaluated.

Methods: Eighty-four individuals referred for lung cancer screening (mean age: 54.5±10.8 years) underwent low-dose computed tomography (LDCT) and ultra-low-dose CT. CT examinations were performed with attenuation-based tube current modulation. Reference tube voltage and current were set to 120kV/25mÅs for LDCT and 80kV/4mÅs for ultra-low-dose CT. CT images were reconstructed with filtered back projection (FBP) for LDCT, and with FBP and IR for ultra-low-dose CT datasets. A reference standard was established by a consensus panel of 2 different radiologists on LDCT. Volume and diameter of the solid nodules were measured on LDCT with FBP and ultra-low dose CT with FBP and IR. Interobserver and interscan variability were analyzed and compared by the Bland-Altman method.

Results: A total of 127 nodules were identified, including 105 solid nodules, 15 part solid nodules, 7 ground glass nodules. On ultra-low-dose CT scans, the effective radiation dose was 0.13±0.11mSv. A total of 113 (88.9%) and 110 (86.6%) true-positive nodules with FBP versus 117 (92.1%) and 118(92.9%) with IR were detected by two observers, respectively. The volume and size of the 105 solid nodules were measured, with mean volume/diameter of 46.5±46.6 mm(3)/5.1±1.6mm. There was no significant difference in nodule volume or diameter measurements between ultra-low-dose CT and LDCT protocols for solid nodules.

Conclusions: Ultra-low-dose CT with iterative reconstruction has high sensitivity for lung nodule detection without significant difference in nodule size and volume measurement compared to LDCT.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ejrad.2015.12.013DOI Listing

Publication Analysis

Top Keywords

solid nodules
16
iterative reconstruction
12
nodules
9
pulmonary nodules
8
ultra-low-dose
8
nodules ultra-low-dose
8
ultra-low-dose iterative
8
computed tomography
8
ldct fbp
8
volume diameter
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!