Accuracy of Intravenous Electrocardiography Confirmation of Peripherally Inserted Central Catheter for Parenteral Nutrition.

Nutr Clin Pract

Division of Endocrinology, Diabetes, Metabolism and Nutrition, Mayo Clinic, Rochester, Minnesota, USA Division of General Internal Medicine, Mayo Clinic, Rochester, Minnesota, USA Division of Gastroenterology and Hepatology, Mayo Clinic, Rochester, Minnesota, USA Division of Gastroenterology, Hepatology and Nutrition, University of Louisville, Kentucky.

Published: April 2016

Background: Parenteral nutrition (PN) is a life-saving therapy for patients with intestinal failure. Safe delivery of hyperosmotic solution requires a central venous catheter (CVC) with tip in the lower superior vena cava (SVC) or at the SVC-right atrium (RA) junction. To reduce cost and delay in use of CVC, new techniques such as intravascular electrocardiogram (ECG) are being used for tip confirmation in place of chest x-ray (CXR). The present study assessed for accuracy of ECG confirmation in home PN (HPN).

Methods: Records for all patients consulted for HPN from December 17, 2014, to June 16, 2015, were reviewed for patient demographics, diagnosis leading to HPN initiation, and ECG and CXR confirmation. CXRs were subsequently reviewed by a radiologist to reassess location of the CVC tip and identify those that should be adjusted.

Results: Seventy-three patients were eligible, and after assessment for research authorization and postplacement CXR, 17 patients (30% male) with an age of 54 ± 14 years were reviewed. In all patients, postplacement intravascular ECG reading stated tip in the SVC. However, based on CXR, the location of the catheter tip was satisfactory (low SVC or SVC-RA junction) in 10 of 17 patients (59%).

Conclusion: Due to the high osmolality of PN, CVC tip location is of paramount importance. After radiology review of CXR, we noted that 7 of 17 (41%) peripherally inserted central catheter lines were in an unsatisfactory position despite ECG confirmation. With current data available, intravenous ECG confirmation should not be used as the sole source of tip confirmation in patients receiving HPN.

Download full-text PDF

Source
http://dx.doi.org/10.1177/0884533615621548DOI Listing

Publication Analysis

Top Keywords

ecg confirmation
16
peripherally inserted
8
inserted central
8
central catheter
8
parenteral nutrition
8
confirmation
7
patients
7
ecg
6
cxr
5
accuracy intravenous
4

Similar Publications

Police officers are exposed to high levels of stress. Serving on Special Weapons and Tactics (SWAT) teams is a highly demanding duty that may further increase levels of stress in police personnel. This stress may accumulate, thereby increasing allostatic load.

View Article and Find Full Text PDF

Cardiac strangulation with chronic ab-extrinseco occlusion of the left-circumflex artery from an epicardial lead: a case report.

Eur Heart J Case Rep

January 2025

Department of Radiological and Hematological Sciences, Section of Radiology, Università Cattolica del Sacro Cuore, Largo Agostino Gemelli 8 - 00168 Rome, Italy.

Background: Cardiac strangulation (CS) from epicardial pacing leads (EPLs) is a rare and potentially lethal mechanical complication associated with epicardial pacemaker (PM) implantation.

Case Summary: We report a case of a 44-year-old-female patient presenting with chest and left shoulder pain in the absence of reported trauma with history of congenital atrioventricular block treated with epicardial PM implantation during the childhood and subsequent transvenous reimplantation over the years. Troponin I resulted within normal values and ECG, transthoracic echocardiography and chest X-ray documented no acute cardiopulmonary findings.

View Article and Find Full Text PDF

Background: Individuals with spinal cord injury (SCI) commonly have autonomic dysreflexia (AD) with increased sympathetic activity. After SCI, individuals have decreased baroreflex sensitivity and increased vascular responsiveness.

Objective: To evaluate the relationship between baroreflex and blood vessel sensitivity with AD symptoms.

View Article and Find Full Text PDF

An atrial septal defect (ASD) is a common congenital heart anomaly that results in irregular blood flow between the systemic and pulmonary circulations due to an opening in the atrial septum. Ostium secondum ASD accounts for a large proportion of these defects and often goes unnoticed during childhood and adolescence. Pulmonary hypertension (PH), affecting a significant number of patients with ostium secondum ASD, is associated with functional limitations, heart failure, and tachyarrhythmias.

View Article and Find Full Text PDF

Purpose: Resting beat-to-beat blood pressure variability is a strong predictor of cardiovascular events and mortality. However, its underlying mechanisms remain incompletely understood. Given that the sympathetic nervous system plays a pivotal role in cardiovascular regulation, we hypothesized that alpha-1 adrenergic receptors (the main sympathetic receptor controlling peripheral vasoconstriction) may contribute to resting beat-to-beat blood pressure variability.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!