Aims: Active-fixation leads have been associated with higher incidence of cardiac perforation. Large series specifically evaluating this complication are lacking. We sought to evaluate the incidence and predictors of clinically relevant cardiac perforation in a consecutive series of patients implanted with active-fixation pacing and defibrillation leads.

Methods And Results: We conducted a retrospective observational study including all consecutive patients implanted with an active-fixation pacing/defibrillation lead at our institution from July 2008 to July 2015. The incidence of clinically relevant cardiac perforation and cardiac tamponade was evaluated. Univariate and multivariate analyses were used to identify predictors of cardiac perforation. Acute and long-term management of these patients was also investigated. A total of 3822 active-fixation pacing (n = 3035) and defibrillation (n = 787) leads were implanted in 2200 patients. Seventeen patients (0.8%) had clinically relevant cardiac perforation (13 acute and 4 subacute perforations), and 13 (0.5%) had cardiac tamponade resolved with pericardiocentesis. None of the patients with cardiac perforation required surgical treatment. In multivariate analysis, an age >80 years (OR 3.84, 95% CI 1.14-12.87, P = 0.029), female sex (OR 3.14, 95% CI 1.07-9.22, P = 0.037), and an apical position of the right ventricular lead (OR 3.37, 95% CI 1.17-9.67, P = 0.024) were independent predictors of cardiac perforation.

Conclusions: Implantation of active-fixation leads is associated with a low incidence of clinically relevant cardiac perforation. Older and female patients have a higher risk of perforation as well as those patients receiving the ventricular lead in an apical position.

Download full-text PDF

Source
http://dx.doi.org/10.1093/europace/euv410DOI Listing

Publication Analysis

Top Keywords

cardiac perforation
32
clinically relevant
20
relevant cardiac
20
active-fixation pacing
12
cardiac
11
perforation
9
incidence predictors
8
predictors clinically
8
implantation active-fixation
8
pacing defibrillation
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!