Background: Serum immunoglobulin free light chains (FLC) are secreted into circulation by plasma cells as a by-product of immunoglobulin production. In a healthy individual the population of FLC is polyclonal as no single cell is secreting more FLC than the total immunoglobulin secreting cell population. In a person with a plasma cell dyscrasia, such as multiple myeloma (MM) or light chain amyloidosis (AL), a clonal population of plasma cells secretes a monoclonal light chain at a concentration above the normal polyclonal background.
Methods: We recently showed that monoclonal immunoglobulin rapid accurate mass measurement (miRAMM) can be used to identify and quantify a monoclonal light chain (LC) in serum and urine above the polyclonal background. This was accomplished by reducing immunoglobulin disulfide bonds releasing the LC to be analyzed by microLC-ESI-Q-TOF mass spectrometry. Here we demonstrate that the methodology can also be applied to the detection and quantification of FLC by analyzing a non-reduced sample.
Results: Proof of concept experiments were performed using purified FLC spiked into normal serum to assess linearity and precision. In addition, a cohort of 27 patients with AL was analyzed and miRAMM was able to detect a monoclonal FLC in 23 of the 27 patients that had abnormal FLC values by immunonephelometry.
Conclusions: The high resolution and high mass measurement accuracy provided by the mass spectrometry based methodology eliminates the need for κ/λ ratios as the method can quantitatively monitor the abundance of the κ and λ polyclonal background at the same time it measures the monoclonal FLC.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1515/cclm-2015-0917 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!