Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Goal: In refractive surgery, astigmatism-correcting treatments are generally planned with the aid of some diagnostic imaging device and often executed by some computer guided laser system. In the transition from sitting down at a diagnostic device to lying down beneath a laser system, a phenomenon known as cyclotorsion (rotation of the eye within the socket) occurs. Hence, registration between lasers and diagnostic devices is necessary. The purpose of this paper is to present a newly developed algorithm that accomplishes robust registration using images of the patient's iris in the context of laser-assisted cataract surgery, and evaluate its efficacy.
Methods: The proposed iris registration algorithm was tested on real cataract patient images obtained from commercially available devices. Accuracy was measured against manual registrations performed by trained humans. Conservative bounds on success and failure rates were computed using novel statistical methods.
Results: The algorithm better approximated the cyclotorsion as averaged over manual measurements from three trained humans than any of the three individual humans, with a 95% tolerance interval of ±1.36(°) . In addition, a success rate ≥ 99.0% was observed for an acceptance threshold setting that allowed for a false registration rate ≤ 1.00*10(-3)%.
Conclusion: The proposed iris registration algorithm accurately and consistently compensates for cyclotorsion in laser-assisted cataract surgery.
Significance: This paper details the first algorithm to be used for iris registration in laser-assisted cataract surgery. Enabling surgeons to make use of this algorithm in real surgeries is expected to have a significant impact on astigmatism management in cataract surgery.