We present and evaluate an approach for coupling liquid chromatography in glass chips with mass spectrometry via fully integrated electrospray emitters. We developed an instrumental platform which allows a robust and reproducible operation of high performance chip chromatography coupled to mass spectrometry. A comparison of differently shaped emitters, from flat over edged to pulled geometries, revealed that all types performed equally well for typical nano-HPLC flow rates. At very low flow rates below 50 nL·min(-1) very sharp, pulled nanospray emitters turned out to be mandatory for the generation of a stable electrospray.