Background: Stair-stepping motion is important in daily living, similar to gait. Knee prostheses need to have even more superior performance and stability in stair-stepping motion than in gait. The purpose of this analysis was to estimate in vivo knee motion in stair stepping and determine if this unique knee prosthesis function as designed.

Methods: A total of 20 patients with Bi-Surface posterior-stabilizing (PS) implants were assessed. The Bi-Surface PS knee is a posterior-cruciate substitute prosthesis with a unique ball-and-socket joint in the mid-posterior portion of the femoral and tibial components. Patients were examined during stair-stepping motion using a 2-dimensional to 3-dimensional registration technique.

Results: The kinematic pattern in step up was a medial pivot, in which the level of anteroposterior translation was very small. In step down, the kinematic pattern was neither a pivot shift nor a rollback. From minimum to maximum flexion, anterior femoral translation occurred slightly.

Conclusions: In this study, this unique implant had good joint stability during stair stepping. The joint's stability during stair stepping was affected by the design of the femorotibial joint rather than post/cam engagement or the ball-and-socket joint.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4736098PMC
http://dx.doi.org/10.1186/s13018-016-0354-5DOI Listing

Publication Analysis

Top Keywords

stair-stepping motion
12
stair stepping
12
ball-and-socket joint
8
kinematic pattern
8
stability stair
8
knee
5
vivo kinematics
4
unique
4
kinematics unique
4
unique posterior-stabilized
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!