Scanning electron microscopy of chronically implanted intracortical microelectrode arrays in non-human primates.

J Neural Eng

Department of Neurological Surgery, New Jersey Medical School, Rutgers University, Newark, NJ, USA. Department of Neuroscience and Brown Institute for Brain Science, Brown University, Providence, RI, USA.

Published: April 2016

Objective: Signal attenuation is a major problem facing intracortical sensors for chronic neuroprosthetic applications. Many studies suggest that failure is due to gliosis around the electrode tips, however, mechanical and material causes of failure are often overlooked. The purpose of this study was to investigate the factors contributing to progressive signal decline by using scanning electron microscopy (SEM) to visualize structural changes in chronically implanted arrays and histology to examine the tissue response at corresponding implant sites.

Approach: We examined eight chronically implanted intracortical microelectrode arrays (MEAs) explanted from non-human primates at times ranging from 37 to 1051 days post-implant. We used SEM, in vivo neural recordings, and histology (GFAP, Iba-1, NeuN). Three MEAs that were never implanted were also imaged as controls.

Main Results: SEM revealed progressive corrosion of the platinum electrode tips and changes to the underlying silicon. The parylene insulation was prone to cracking and delamination, and in some instances the silicone elastomer also delaminated from the edges of the MEA. Substantial tissue encapsulation was observed and was often seen growing into defects in the platinum and parylene. These material defects became more common as the time in vivo increased. Histology at 37 and 1051 days post-implant showed gliosis, disruption of normal cortical architecture with minimal neuronal loss, and high Iba-1 reactivity, especially within the arachnoid and dura. Electrode tracts were either absent or barely visible in the cortex at 1051 days, but were seen in the fibrotic encapsulation material suggesting that the MEAs were lifted out of the brain. Neural recordings showed a progressive drop in impedance, signal amplitude, and viable channels over time.

Significance: These results provide evidence that signal loss in MEAs is truly multifactorial. Gliosis occurs in the first few months after implantation but does not prevent useful recordings for several years. Progressive meningeal fibrosis encapsulates and lifts MEAs out of the cortex while ongoing foreign body reactions lead to progressive degradation of the materials. Long-term impedance drops are due to the corrosion of platinum, cracking and delamination of parylene, and delamination of silicone elastomer. Oxygen radicals released by cells of the immune system likely mediate the degradation of these materials. Future MEA designs must address these problems through more durable insulation materials, more inert electrode alloys, and pharmacologic suppression of fibroblasts and leukocytes.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4854331PMC
http://dx.doi.org/10.1088/1741-2560/13/2/026003DOI Listing

Publication Analysis

Top Keywords

chronically implanted
12
1051 days
12
scanning electron
8
electron microscopy
8
implanted intracortical
8
intracortical microelectrode
8
microelectrode arrays
8
non-human primates
8
electrode tips
8
days post-implant
8

Similar Publications

Prognostic role of aetiological agent vs. clinical pattern in candidates to lead extraction for cardiac implantable electronic device infections.

Sci Rep

December 2024

Department of Medical and Surgical Sciences, Institute of Cardiology, University of Bologna, Policlinico S.Orsola-Malpighi, via Massarenti 9, Bologna, 40138, Italy.

Cardiac implantable electronic devices infections (CIEDI) are associated with poor survival despite the improvement in transvenous lead extraction (TLE). Aetiology and systemic involvement are driving factors of clinical outcomes. The aim of this study was to explore their contribute on overall mortality.

View Article and Find Full Text PDF

BAY 2413555 is a novel selective and reversible positive allosteric modulator of the type 2 muscarinic acetylcholine (M2) receptor, aimed at enhancing parasympathetic signaling and restoring cardiac autonomic balance for the treatment of heart failure (HF). This study tested the safety, tolerability and pharmacokinetics of this novel therapeutic option. REMOTE-HF was a multicenter, double-blind, randomized, placebo-controlled, phase Ib dose-titration study with two active arms.

View Article and Find Full Text PDF

Background: Distal radius fractures are prevalent in small and toy-breed dogs, presenting significant treatment challenges due to complications such as delayed union or non-union. These complications are often exacerbated by reduced vascular density at the distal diaphyseal-metaphyseal junction of the radius, which is vital for bone healing, particularly in toy and small breed dogs. Circular external fixation (CEF) is known for its effectiveness in managing acute and chronic fractures and providing temporary stabilization in humans.

View Article and Find Full Text PDF

Background: Treating infectious bone defects combined with large soft-tissue lesions poses significant clinical challenges. Herein, we introduced a modified two-stage treatment approach involving the implantation of 3D-printed prostheses and flap repair to treat large segmental infectious tibial bone defects.

Method: We conducted a retrospective study of 13 patients treated at our center between April 2018 and March 2022 for tibial infections owing to posttraumatic infection and chronic osteomyelitis combined with soft tissue defects.

View Article and Find Full Text PDF

Objective: To present the clinical result of spinal fixation system made entirely of Carbon-Fiber-Reinforced (CFR)-Hybrid Polyaryl-Ether-Ether-Ketone (PEEK).

Summary Of Background Data: Fusion surgery has been used to treat chronic low back pain caused by degenerative disk disease (DDD). The traditional pedicle screw system made of titanium, though biocompatible, can lead to complications, such as stress shielding and implant failure.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!