Objective: The purpose of this article is to investigate the magnitude of dose optimization for a manufacturer-recommended urolithiasis protocol in a second-generation dual-source CT scanner.
Materials And Methods: Custom renal phantoms with 24 stones were scanned using the manufacturer-provided dual-energy CT protocol (tube A, 100 kVp and 210 reference mAs; tube B, 140 kVp and 162 reference mAs) and seven dose-optimized protocols in which the reference tube current-time product setting of tube A was reduced stepwise by 20 mAs. Detection and characterization of the stones was assessed. In the patient study, 25 patients underwent the manufacturer-provided dual-energy protocol and 25 patients underwent imaging with a dose-optimized protocol (tube A, 100 kVp and 90 reference mAs; tube B, 140 kVp and 70 reference mAs). Dose-length product (DLP), image noise, and contrast-to-noise ratio (CNR) were assessed. Subjective image quality was analyzed by three independent radiologists.
Results: In the phantom study, the reference tube current-time product of tube A could be reduced from 210 to 90 mAs without losing the accuracy of detection or characterization of the calculi. In the patient study, the dose-optimized protocol resulted in a significant reduction of the average DLP by 51% compared with the standard protocol (219.4 vs 443.5 mGy·cm, respectively; p = 0.0001). The image noise was higher, and the CNR was lower, in the dose-optimized group than in the standard-dose group (p < 0.05). The subjective overall image quality of the dose-optimized CT examinations was rated as good, and that of the standard-dose CT examinations was rated as excellent (p = 0.001).
Conclusion: The in vitro and in vivo assessment revealed a potential for a 51% dose reduction of the manufacturer-recommended dual-energy CT protocol for urolithiasis without compromising the accuracy.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.2214/AJR.15.15204 | DOI Listing |
Nutrients
December 2024
Department of Anesthesiology, Reanimatology and Intensive Care, University Hospital Split, 21000 Split, Croatia.
Background: Studies have shown a high prevalence of vitamin D deficiency in critically ill patients, and these patients are at higher risk for pneumonia and have increased incidence of sepsis and mortality. In this study, we reviewed available literature from randomized controlled trials (RCTs) on vitamin D supplementation in critically ill patients and summarized the evidence in this narrative review.
Methods: Randomized controlled trials that included vitamin D supplementation as an intervention were eligible for inclusion.
Diagnostics (Basel)
January 2025
Faculty of Applied Science, Uplands College of Science and Technology Incorporated (UCSI), No. 1, Jalan Menara Gading, Kuala Lumpur 56000, Malaysia.
: This study compares system-reported organ doses (ODs) to manually calculated mean glandular doses (MGDs) in mammography across multiple centers and manufacturers in Dubai. : A retrospective study of 2754 anonymized mammograms from six clinics in Dubai were randomly retrieved from a central dose survey database. Organ doses were documented along with other dosimetry information like kVp, mAs, filter, target, compression force, and breast thickness.
View Article and Find Full Text PDFInsights Imaging
January 2025
Department of Medical Imaging Center, Nanfang Hospital, Southern Medical University, Guangzhou, 510515, China.
Objectives: To investigate the image quality and diagnostic performance with ultra-low dose dual-layer detector spectral CT (DLSCT) by various reconstruction techniques for evaluation of pulmonary nodules.
Materials And Methods: Between April 2023 and December 2023, patients with suspected pulmonary nodules were prospectively enrolled and underwent regular-dose chest CT (RDCT; 120 kVp/automatic tube current) and ultra-low dose CT (ULDCT; 100 kVp/10 mAs) on a DLSCT scanner. ULDCT was reconstructed with hybrid iterative reconstruction (HIR), electron density map (EDM), and virtual monoenergetic images at 40 keV and 70 keV.
Clin Kidney J
January 2025
Instituto de Investigación Sanitaria del Principado de Asturias (ISPA), Bone and Mineral Research Unit, REDinREN (RD06/0016/1013, RD12/0021/0023 and RD16/0009/0017) and RICORS2040 (RD21/0005/0019) del ISCIII, Oviedo, Spain.
Background: Ageing often affects biomarker production. Yet, clinical/optimal thresholds to guide clinical decisions do not consider this. Serum albumin decreases with age, but hypoalbuminaemia is defined as serum albumin <4.
View Article and Find Full Text PDFDiagnostics (Basel)
December 2024
Medical Diagnostic Imaging Department, College of Health Science, University of Sharjah, Sharjah 27272, United Arab Emirates.
: Computed Tomography (CT) is crucial in medical diagnosis, particularly for head examinations. Diagnostic Reference Levels (DRLs) are pivotal in balancing diagnostic efficacy with radiation safety. International organizations such as the International Atomic Energy Agency (IAEA) and the International Commission on Radiological Protection (ICRP) provide guidelines for establishing DRLs, emphasizing their importance in optimizing radiation doses.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!