In preliminary ELISA studies where released-active forms (RAF) of antibodies (Abs) to interferon-gamma (IFNg) were added to the antigen-antibody system, a statistically significant difference in absorbance signals obtained in their presence in comparison to placebo was observed. A piezoelectric immunosensor assay was developed to support these data and investigate the effects of RAF Abs to IFNg on the specific interaction between Abs to IFNg and IFNg. The experimental conditions were designed and optimal electrode coating, detection circumstances and suitable chaotropic agents for electrode regeneration were selected. The developed technique was found to provide high repeatability, intermediate precision and specificity. The difference between the analytical signals of RAF Ab samples and those of the placebo was up to 50.8%, whereas the difference between non-specific controls and the placebo was within 5%-6%. Thus, the piezoelectric immunosensor as well as ELISA has the potential to be used for detecting the effects of RAF Abs to IFNg on the antigen-antibody interaction, which might be the result of RAF's ability to modify the affinity of IFNg to specific/related Abs.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4732129PMC
http://dx.doi.org/10.3390/s16010096DOI Listing

Publication Analysis

Top Keywords

abs ifng
12
released-active forms
8
ifng antigen-antibody
8
piezoelectric immunosensor
8
effects raf
8
raf abs
8
ifng
6
abs
5
piezoelectric immunosensors
4
immunosensors detection
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!