Aims: To review the outcomes of studies and the safety of newer transcatheter aortic valves (THV).

Methods And Results: All studies reporting on second-generation THV were identified and pooled using the systematic review guidelines. Twenty-four reports on 1,708 patients and eight THV were included in the analysis. The pooled 30-day event rate for mortality after transcatheter aortic valve implantation (TAVI) was 5.7% (95% CI: 4.0-7.8), myocardial infarction (MI) was 1.7% (95% CI: 1.1-2.6), stage 3 acute kidney injury (AKI) was 3.4% (95% CI: 2.0-5.6), life-threatening bleeding was 5.1% (95% CI: 3.3-7.8), major vascular complications was 4.9% (95% CI: 3.5-6.6%), major bleeding was 10.5% (95% CI: 5.1-20.4), major stroke was 2.4% (95% CI: 1.7-3.4), permanent pacemaker utilisation was 13.5% (95% CI: 10.8-16.9), and coronary obstruction was 1.2% (95% CI: 0.6%-2.4%). Moderate or severe aortic insufficiency (AI) after TAVI was 4.2% (95% CI: 2.0-8.5). The pooled 30-day mean gradient and effective orifice area (EOA) were 11.63 mmHg (95% CI: 10.19-13.07) and 1.60 cm2 (95% CI: 1.5-1.7), respectively. All estimates compare favourably to events reported for first-generation valves.

Conclusions: Our findings suggest that the new THV have a low risk of TAVI-related short-term complications.

Download full-text PDF

Source
http://dx.doi.org/10.4244/EIJV11I9A211DOI Listing

Publication Analysis

Top Keywords

transcatheter aortic
12
95%
12
systematic review
8
aortic valves
8
pooled 30-day
8
review safety
4
safety second-generation
4
second-generation transcatheter
4
aortic
4
valves aims
4

Similar Publications

Transcatheter versus Surgical Aortic Valve Replacement in Patients Younger than 65 Years in the US.

J Thorac Cardiovasc Surg

January 2025

Department of Cardiac Surgery, Smidt Heart Institute, Cedars-Sinai Medical Center, Los Angeles, California, USA. Electronic address:

Objective: To characterize trends and outcomes of aortic valve replacement in patients <65 with aortic stenosis between 2013 and 2021.

Methods: This retrospective analysis included 9,557 patients who underwent biological aortic valve replacement in California, New York, and New Jersey from 2013 through 2021. Patients were stratified by approach: transcatheter aortic valve replacement (TAVR) versus surgical aortic valve replacement (SAVR).

View Article and Find Full Text PDF

Background: Evidence regarding the incidence of prosthesis-patient mismatch (PPM) and long-term mortality after transcatheter aortic valve replacement (TAVR) in patients with bicuspid aortic valve stenosis (AS) is scarce.

Objectives: This study sought to assess the incidence and prognostic impact of PPM after TAVR for bicuspid AS compared with that for tricuspid AS.

Methods: In total, 7,393 patients who underwent TAVR were prospectively enrolled in the OCEAN-TAVI (Optimized Catheter Valvular Intervention Transcatheter Aortic Valve Implantation) registry, an ongoing Japanese, multicenter registry.

View Article and Find Full Text PDF

Severe mitral regurgitation (MR) following surgical repair of the mitral valve poses a significant clinical challenge. Patients who have undergone surgery are typically at high risk for a second operation. This report details the case of a 54-year-old male who underwent aortic valve replacement and mitral valve repair using a 34-ring, 14 years prior.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!