Objectives: The effects of combining multiple calibrations on assay accuracy (bias) and measurement of calibration stability were investigated for total triiodothyronine (TT3), vitamin B12 and luteinizing hormone (LH) using Beckman Coulter's Access 2 analyzer.
Methods: Three calibration procedures (CC1, CC2 and CC3) combined 12, 34 and 56 calibrator measurements over 1, 2, and 3 days. Bias was calculated between target values and average measured value over 3 consecutive days after calibration. Using regression analysis of calibrator measurements versus measurement date, calibration stability was determined as the maximum number of days before a calibrator measurement exceeded 5% tolerance limits.
Results: Competitive assays (TT3, vitamin B12) had positive time regression slopes, while sandwich assay (LH) had a negative slope. Bias values for TT3 were -2.49%, 1.49%, and -0.50% using CC1, CC2 and CC3 respectively, with calibrator stability of 32, 20, and 30 days. Bias values for vitamin B12 were 2.44%, 0.91%, and -0.50%, with calibrator stability of 4, 9, and 12 days. Bias values for LH were 2.26%, 1.44% and -0.29% with calibrator stability of >43, 39 and 36 days. Measured stability was more consistent across calibration procedures using percent change rather than difference from target: 26 days for TT3, 12 days for B12 and 31 days for LH.
Conclusions: Averaging over multiple calibrations produced smaller bias, consistent with improved accuracy. Time regression slopes in percent change were unaffected by number of calibration measurements but calibrator stability measured from the target value was highly affected by the calibrator value at time zero.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.3109/00365513.2015.1126760 | DOI Listing |
Metrologia
January 2024
National Institute of Standards and Technology, 100 Bureau Dr, Gaithersburg, MD 20899, United States of America.
Upgrades to the vacuum wavelength calibration service at the National Institute of Standards and Technology are reported. The instrumentation centerpiece is an optical frequency comb stabilized to a GPS-disciplined oscillator, thereby providing direct traceability to the SI second. Historically, the service has covered lasers at the popular interferometry wavelengths red and green.
View Article and Find Full Text PDFFood Chem X
January 2025
College of Food Science and Engineering, Shanxi Agricultural University, Taiyuan, Shanxi, China.
Aiming to enable online freshness-monitoring of meat within modified-atmosphere package, we developed a ratiometric array that was fluorescently responsive to volatile organic compounds-ammonia (NH) released by protein decaying. The array was consisted of two 3 mm × 6 mm rectangles precisely and uniformly printed with fluorescein isothiocyanate (FITC) as indicator and rhodamine B (RhB) as internal reference on the filter-paper, respectively. The fluorescence intensity of the array area was calibrated according to Green/Red ratio of the digitalized pixels extracted from images facilitated by a smartphone.
View Article and Find Full Text PDFAngew Chem Int Ed Engl
January 2025
Nankai University, School of Materials Science and Engineering, Smart Sensing Interdisciplinary Science Center, Collaborative Innovation Center of Chemical Science and Engineering, CHINA.
The practical applications of activation-unstable mesoporous metal-organic frameworks (MOFs) are often constrained by their structural instability. However, enhancing their stability could unlock valuable functionalities. Herein, we stabilized the otherwise unstable, post-activated structure of a novel mesoporous Zr(IV)-MOF, NKM-809, which uses a pyridine-containing amphiprotic linker (PPTB).
View Article and Find Full Text PDFBiosens Bioelectron
December 2024
Beijing National Laboratory for Molecular Sciences, College of Chemistry and Molecular Engineering, Peking University, Beijing, 100871, China. Electronic address:
Multi-signal-based self-calibrating biosensors have become a research focus due to their superior accuracy and sensitivity in recent years. Herein, the potential-resolved differential ECL immunoassay based on dual co-reactants regulation was developed. Meso-tetra(4-carboxyphenyl)porphyrin (TCPP) functionalized zirconium dioxide (ZrO) composites (TCPP-ZrO) was first synthesized using TCPP as the luminophore and ZrO as the enhancer and stabilizer.
View Article and Find Full Text PDFBMJ Open
December 2024
Department of Applied Health Sciences, School of Health Sciences, College of Medicine and Health, University of Birmingham, Birmingham, UK.
Introduction: Ewing sarcoma is a rare paediatric cancer. Currently, there is no way of accurately predicting these patients' survival at diagnosis. Disease type (ie, localised disease, lung/pleuropulmonary metastases and other metastases) is used to guide treatment decisions, with metastatic patients generally having worse outcomes than localised disease patients.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!