2-Aminoacetophenone (AAP) is closely correlated with the appearance of the sensory phenomenon of UTA ("untypical aging off-flavor") in wine. AAP analyses are generally performed by gas chromatography and mass selective detection (GC/MS), when AAP is extracted from wines by liquid-liquid, solid-liquid or solid phase microextraction. Here we present a rapid, selective and sensitive method for the determination of AAP in wine by high-performance thin-layer chromatography with fluorescence detection (HPTLC-FLD). As internal standard, 2-amino-4-methoxyacetophenone was used. Liquid-liquid extraction with t-butyl methyl ether was followed by a basic cleanup of the extracts, which were applied onto HPTLC amino plates developed with methylene chloride/toluene (7+3, v/v) as mobile phase. Dipping the dried plate into hexane-paraffin solution enhanced fluorescence that was scanned at 366/>400nm. Limits of detection and quantitation were determined to be 0.1 and 0.3μgL(-1) wine, respectively, while only AAP concentrations >0.5μgL(-1) result in UTA. Recoveries were near 100% for model, white, rosé and red wines. Thus, the HPTLC-FLD method enables the analysis of AAP in wines clearly below the odor thresholds and represents a rapid and convenient screening alternative to existing GC/MS methods.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.chroma.2015.12.081 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!