Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: Given the degree of variation in clavicular morphology, 4 clavicle plating systems were examined for their congruity as superior, midshaft, anatomic clavicle (SMAC) plates in a cadaveric study.
Methods: SMAC plates from 4 manufacturers were applied to 79 dry right human clavicles. Two systems offered multiple (4) variations of plates (MP), 1 offered two variations (TP), and 1 had a single plate (SP). Two examiners applied and clamped the best-fitting plate from each system onto each of the 79 clavicles and then graded them: 1, poor fit; 2, good fit; and 3, anatomic fit. Each examiner repeated the process to assess intraobserver and interobserver reliability. The scores were averaged to produce a final score for each system for each clavicle.
Results: The MP systems scored the highest (32%-37% anatomic, 54%-63% good, 5%-8% poor), followed by the TP system (30% anatomic, 53% good, 17% poor), and finally the SP system (9% anatomic, 59% good, 32% poor). Of note, clavicular length significantly correlated with a higher degree of conformity in all plating systems (Spearman rank correlation P < .05 for each system). In clavicles longer than 150 mm, the MP and TP systems performed identically, with the SP system close behind. Contouring of the plate is needed in 73% of cases overall.
Conclusion: Plating systems with multiple plate shape variations are more advantageous when dealing with smaller-sized clavicles, typically in females. However, when dealing with larger clavicles, there was no real difference.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jse.2015.10.020 | DOI Listing |