Background: A liquid chromatography-tandem mass spectrometric (LC-MS/MS) method was developed for routine measurement of ribavirin concentrations in EDTA-anticoagulated plasma.

Methods: After protein precipitation, we used a bridged ethylene hybrid (hydrophilic interaction) chromatography column, 0.1 mmol/L ammonium formate pH 3.0, and a gradient of 85%-96% acetonitrile to achieve baseline separation of ribavirin from isobaric uridine. Quantitation was assured using both primary (m/z 245.3 > 113.0) and secondary transitions (m/z 245.3 > 96.0) of the protonated species. Chromatographic separation and column washing also negated interference from major phospholipid species.

Results: There was a linear relationship between concentration and response to 10 mg/L, with a minimum detectable level and a minimum level of quantitation both of 0.1 mg/L. Imprecision within the assay was <10% at 0.1 mg/L and <6% between assays for concentrations >0.4 mg/L. Bias was <4%. In clinical samples (n = 12), there was no difference in ribavirin concentrations obtained by an established liquid chromatographic assay with ultraviolet detection. Ribavirin concentrations were stable in plasma stored at room temperature for 3 days but then decreased significantly on day 7. Plasma concentrations were stable for 15 weeks at -20 °C. Concentrations in plasma separated from whole blood at room temperature fell by a median of 19.4% at 4 hours and then rose substantially (median 251% by 3 days). Dose-normalized ribavirin concentrations reached a steady state after a mean of >6 weeks treatment in 76 patients with hepatitis C.

Conclusions: A hydrophilic interaction liquid chromatography-tandem mass spectrometric method to measure ribavirin in plasma was developed. Samples for ribavirin estimation should be kept at 4 °C, separated within 2 hours of collection and stored at 4 °C before analysis, with long-term storage at -20 °C. This method was applied to a study of the ribavirin therapeutic monitoring in patients with hepatitis C.

Download full-text PDF

Source
http://dx.doi.org/10.1097/FTD.0000000000000232DOI Listing

Publication Analysis

Top Keywords

m/z 2453
8
concentration monitoring
4
monitoring plasma
4
plasma ribavirin
4
ribavirin validation
4
validation liquid
4
liquid chromatography-mass
4
chromatography-mass spectrometric
4
spectrometric method
4
method clinical
4

Similar Publications

Antithrombotic Treatment for Stroke Prevention in Cervical Artery Dissection: The STOP-CAD Study.

Stroke

April 2024

Department of Neurology, The Warren Alpert Medical School of Brown University, Brown Medical School, Providence, RI (S. Yaghi, L.S., D. Mandel, K.P., V.D., K.B., T.B., N.K., F. Khan, C.S., N.M., E.G., K.F.).

Background: Small, randomized trials of patients with cervical artery dissection showed conflicting results regarding optimal stroke prevention strategies. We aimed to compare outcomes in patients with cervical artery dissection treated with antiplatelets versus anticoagulation.

Methods: This is a multicenter observational retrospective international study (16 countries, 63 sites) that included patients with cervical artery dissection without major trauma.

View Article and Find Full Text PDF

This study describes the simultaneous detection of positively and negatively charged microparticles by ion trap mass spectrometry (IT-MS) as a novel analytical measurement technique. The instrument was configured with a feeding capillary for particle introduction, an ion trap, and a charge detector that responds to both ions simultaneously. Positively and negatively charged particles are generated by the triboelectric effect inside the capillary entrance of the instrument.

View Article and Find Full Text PDF

Development of a focused high-energy macromolecular ion beam.

Analyst

May 2021

Genomics Research Center, Academia Sinica, Taipei 11529, Taiwan. and Department of Chemistry, National Taiwan University, Taipei 10617, Taiwan.

In this work, we report the development of a focused macromolecular ion beam with kinetic energy of up to 110 keV. The system consists of a quadrupole ion trap (QIT), einzel lens and linear accelerator (LINAC). Based on the combination of matrix-assisted laser desorption ionization (MALDI) and quadrupole ion trapping (QIT), ions were desorbed from the surface and trapped with an ion trap to form biomolecular ion packets.

View Article and Find Full Text PDF

A liquid chromatography-electrospray ionization-tandem mass spectrometry (LC-ESI-MS/MS) method for the quantification of the opioid peptide DAMGO in rat plasma, as well as DAMGO and the microdialysis recovery calibrator [(13)C(2),(15)N]-DAMGO in microdialysis samples, is described. The microdialysis samples consisted of 15 μL Ringer solution containing 0.5% bovine serum albumin.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!