Riata and Riata ST implantable cardioverter-defibrillator leads are prone to structural and electrical failure (EF). Our objective was to evaluate Riata/ST lead performance over a long-term follow-up. Of 184 patients having undergone Riata/ST and Riata ST Optim lead implantation from September 2003 to June 2008, 154 patients were evaluated for EF and radiographic conductor externalization (CE). Survival analysis for EF was performed for Riata/ST leads, both for failure-free lead survival and cumulative hazard. Subanalysis on 7Fr leads was performed to evaluate EF and CE rates both for different Riata ST lead management (monitoring vs proactive) and between Riata ST and Riata ST Optim leads. During a mean follow-up of 7 years, Riata/ST lead EF rate was 13% overall. Similar failure-free survival rate was noted for 7Fr as for 8Fr leads (log-rank, p = 0.63). Of all failed leads, 64% failed only after 5 years of follow-up. Compared with the absolute failure rate of 1.84% per device year, cumulative hazard analysis for leads surviving past 5 years revealed an estimated failure rate of 7% per year. No clinical or procedural predictors for EF were found. The subanalysis on 7Fr leads showed an excellent outcome both for a proactive lead management approach as for Optim leads. In conclusion, long-term survival of the Riata/ST lead is impaired with an accelerating EF risk over time. An initial exponential trend was followed by a linear lead failure pattern for leads surviving past 5 years, corresponding to an estimated 7% annual EF rate. These findings may have repercussions on the lead management strategy in patients currently surviving with a Riata/ST lead to prevent significant clinical events like inappropriate shocks or failed device interventions.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.amjcard.2015.12.013DOI Listing

Publication Analysis

Top Keywords

riata/st lead
16
lead management
12
lead
11
leads
10
implantable cardioverter-defibrillator
8
riata riata
8
riata optim
8
cumulative hazard
8
subanalysis 7fr
8
7fr leads
8

Similar Publications

Background: The Cardiac Lead Assessment Study (CLAS) was a large prospective, multicenter, international postmarket surveillance study conducted at 45 sites.

Objective: The purpose of CLAS was to examine the prevalence and incidence of externalized conductors and electrical dysfunction in subjects with selected St. Jude Medical defibrillator and left ventricular leads.

View Article and Find Full Text PDF

Background: St Jude Medical (now Abbott) Optim-insulated implantable cardioverter-defibrillator (ICD) leads were expected to overcome problems with insulation abrasion and externalized conductors in earlier models. Long-term follow-up is essential to the evaluation of lead performance.

Objective: To determine, in a prospective cohort of Optim-insulated ICD leads, the rates of all-cause mechanical failure and its subtypes (conductor fracture, insulation abrasion, externalized conductors, and other mechanical failures) and electrical dysfunction adjudicated as nonmechanical failure.

View Article and Find Full Text PDF

Background: The Medtronic Sprint Fidelis® and Abbott Riata®/Riata ST® leads are at risk of failure and are subject to FDA recall. Comparative risks of various lead management strategies during elective generator change in a multi-center population are unknown. We aim to describe patients with functional, recalled ICD leads undergoing elective generator replacement and report outcomes according to lead management strategies.

View Article and Find Full Text PDF

A 17-year-old male with hypertrophic cardiomyopathy underwent placement of a dual-chamber implantable cardioverter defibrillator (ICD) in 2010. In October 2016, he suffered a cardiac arrest. His ICD was interrogated after visiting our hospital.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!