The aim of this study was to evaluate the influence of the physical assessment of different light-curing units from 55 dental offices on the irradiance and composite microhardness top/bottom ratio, and the influence of the radiometers for LED or QTH light sources on irradiance measurement. The irradiance of each light-curing unit was evaluated with two radiometers, either for LED or QTH light. A questionnaire regarding the type of source (LED or QTH), time of use, date of last maintenance and light-curing performance assessment applied. The physical assessments were evaluated regarding damage or debris on the light tip. For each light-curing unit, three composite specimens were made (diameter = 7 mm; thickness = 2 mm) with polymerizing time of 20 s, in order to perform the microhardness (Knoop) test. Data were analyzed by Kruskal-Wallis and Dunn test (α = 0.01). There was wide variation in irradiance (0-1000 mW/cm(2)). Approximately 50 % of the light-curing units presented radiation lower than 300 mW/cm(2); 10 % of light-curing units, especially those with LED source, presented values higher than 800 mW/cm(2), and 43 % of light-curing units worked with adequate irradiance between 301 and 800 mW/cm(2). In almost 60 % of cases, no maintenance of light-curing units was performed in a period of 3 to 10 years. The age of the light-curing units and the use of inadequate tips interfered negatively in irradiance. The data emphasize the importance of periodic maintenance of light-polymerizing, light-curing units.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s10266-015-0229-yDOI Listing

Publication Analysis

Top Keywords

light-curing units
32
led qth
12
light-curing
11
influence physical
8
physical assessment
8
assessment light-curing
8
units
8
irradiance composite
8
composite microhardness
8
microhardness top/bottom
8

Similar Publications

Aim Dental light-curing units (LCUs) are used at Matsumoto Dental University Hospital (MDUH). However, the time of installation of the same type of light irradiator varies, which affects the output intensity of the LCU. The purpose of this study was to evaluate the performance of LCU with different output intensities by comparing the depth of cure (DOC) of resin-based composites (RBCs).

View Article and Find Full Text PDF

Light curing infection control barriers: do some types jeopardize the concept of conventional bulk-fill composites?

BMC Oral Health

November 2024

Biomaterials Department, Faculty of Dentistry, Ain-Shams University, Organization of African unity, Cairo, Egypt.

Background: Using infection control barriers (ICBs) on light curing units (LCUs) became mandatory to achieve proper infection control measures without jeopardizing the integrity of the restorations, especially at deeper layers. This study explored the effect of two ICBs on the irradiance of the LCU, as well as the degree of conversion (DC) and flexural strength (FS) of two types of bulk-fill composites. Water vapor permeability (WVP) of both barriers was also assessed to evaluate the capability of such barriers to prevent transmission of blood and saliva droplets and aerosols.

View Article and Find Full Text PDF

Impact of Battery Levels of a Cordless LED Curing Unit on Resin Cement under Varied Lithium Disilicate Thicknesses and Translucencies.

Oper Dent

November 2024

*Luís Henrique Araújo Raposo, DDS, MSc, PhD, associate professor, Department of Occlusion, Fixed Prosthodontics and Dental Materials, School of Dentistry, Federal University of Uberlândia, MG, Brazil.

Objectives: This study aimed to evaluate the impact of battery levels on the emission of a multi-peak cordless LED light-curing unit (LCU) and the effect on the degree of conversion (DC) and Knoop hardness (KH) of a light-cure resin luting agent activated through varying lithium disilicate (LiS2) ceramic thicknesses and translucencies.

Methods: High and low translucency LiS2 discs (IPS e.max Press HT and LT, respectively; shade A1) with thickness of 0.

View Article and Find Full Text PDF

Objective: To evaluate the influence of battery level on power (mW), emission spectrum (mW/cm2/ nm), and light distribution on the active tip (mW/ cm2) of certified (FDA/ANVISA) and low-cost uncertified light-curing units (LCUs) purchased through e-commerce.

Methods: Seven LCUs, three certified: VALO Grand (Ultradent); Radii Xpert (SDI); and LED.B (Woodpecker); and four uncertified: 1 Sec; BS 300; LED curing light; and VAFU (VRN, AZDENT), were used.

View Article and Find Full Text PDF

Influence of light-polymerizing units and zirconia on the physical, chemical and biological properties of self-adhesive resin cements.

BMC Oral Health

October 2024

Center of Excellence in Genomics and Precision Dentistry, Department of Physiology, Faculty of Dentistry, Chulalongkorn University, Bangkok, 10330, Thailand.

Article Synopsis
  • * Results show that using the Kerr DemiPlus LPU with a zirconia screen decreased the hardness and increased the roughness of the cements while also leading to higher levels of residual monomers and negatively impacting hGFs' health and function.
  • * The findings suggest that care in selecting both the materials and polymerization techniques can significantly reduce potential harm to surrounding gum tissues during clinical dental procedures.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!