The objective of this study was to investigate the optimization of a high energy in-line phase sensitive x-ray imaging prototype under different geometric and operating conditions for mammography application. A phase retrieval algorithm based on phase attenuation duality (PAD) was applied to the phase contrast images acquired by the prototype. Imaging performance was investigated at four magnification values of 1.67, 2, 2.5 and 3 using an acrylic edge, an American College of Radiology (ACR) mammography phantom and contrast detail (CD) phantom with tube potentials of 100, 120 and 140 kVp. The ACR and CD images were acquired at the same mean glandular dose (MGD) of 1.29 mGy with a computed radiography (CR) detector of 43.75 μm pixel pitch at a fixed source to image distance (SID) of 170 cm. The x-ray tube focal spot size was kept constant as 7 μm while a 2.5 mm thick aluminum (Al) filter was used for beam hardening. The performance of phase contrast and phase retrieved images were compared with computer simulations based on the relative phase contrast factor (RPF) at high x-ray energies. The imaging results showed that the x-ray tube operated at 100 kVp under the magnification of 2.5 exhibits superior imaging performance which is in accordance to the computer simulations. As compared to the phase contrast images, the phase retrieved images of the ACR and CD phantoms demonstrated improved imaging contrast and target discrimination. We compared the CD phantom images acquired in conventional contact mode with and without the anti-scatter grid using the same prototype at 1.295 mGy and 2.59 mGy using 40 kVp, a 25 μm rhodium (Rh) filter. At the same radiation dose, the phase sensitive images provided improved detection capabilities for both the large and small discs, while compared to the double dose image acquired in conventional mode, the observer study also indicated that the phase sensitive images provided improved detection capabilities for the large discs. This study therefore validates the potential of using high energy phase contrast x-ray imaging to improve lesion detection and reduce radiation dose for clinical applications such as mammography.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4711281PMC
http://dx.doi.org/10.3233/XST-150519DOI Listing

Publication Analysis

Top Keywords

phase contrast
20
phase sensitive
16
phase
13
high energy
12
images acquired
12
in-line phase
8
imaging prototype
8
x-ray imaging
8
images
8
contrast images
8

Similar Publications

Application value of split-bolus contrast injection combined with dual-energy CT scanning technology in pediatric CTU imaging.

Eur J Radiol

January 2025

Department of Radiology Children's Hospital of Chongqing Medical University, National Clinical Research Center for Child Health and Disorders, Ministry of Education Key Laboratory of Child Development and Disorders, Chongqing Key Laboratory of Child Neurodevelopment and Cognitive Disorders, China. Electronic address:

Objective: To explore the clinical value of combining split-bolus contrast injection with dual-energy CT(DECT) scanning technology in pediatric computed tomography urography (CTU) imaging.

Methods: A total of 128 children aged 0-17 years were prospectively selected and randomly assigned to three groups: A, B, and C. For Group A, a high-pitch flash mode was employed, where a single bolus of contrast agent was followed by four-phase scanning (noncontrast, cortex, medulla, and excretory phases).

View Article and Find Full Text PDF

Intraindividual Comparison of Image Quality Between Low-Dose and Ultra-Low-Dose Abdominal CT With Deep Learning Reconstruction and Standard-Dose Abdominal CT Using Dual-Split Scan.

Invest Radiol

January 2025

From the Department of Radiology, Ulsan University Hospital, Ulsan, Republic of Korea (T.Y.L.); Department of Radiology, University of Ulsan College of Medicine, Seoul, Republic of Korea (T.Y.L.); Department of Radiology, Seoul National University Hospital, Seoul, Republic of Korea (J.H.Y., H.K., J.M.L.); Department of Radiology, Seoul National University College of Medicine, Seoul, Republic of Korea (J.H.Y., S.H.P., J.M.L.); Department of Radiology, Inje University Busan Paik Hospital, Busan, Republic of Korea (J.Y.P.); Department of Radiology, Seoul National University Bundang Hospital, Seongnam, Republic of Korea (S.H.P.); Department of Radiology, Hanyang University College of Medicine, Seoul, Republic of Korea (C.L.); Division of Biostatistics, Medical Research Collaborating Center, Seoul National University Hospital, Seoul, Republic of Korea (Y.C.); and Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul, Republic of Korea (J.M.L.).

Objective: The aim of this study was to intraindividually compare the conspicuity of focal liver lesions (FLLs) between low- and ultra-low-dose computed tomography (CT) with deep learning reconstruction (DLR) and standard-dose CT with model-based iterative reconstruction (MBIR) from a single CT using dual-split scan in patients with suspected liver metastasis via a noninferiority design.

Materials And Methods: This prospective study enrolled participants who met the eligibility criteria at 2 tertiary hospitals in South Korea from June 2022 to January 2023. The criteria included (a) being aged between 20 and 85 years and (b) having suspected or known liver metastases.

View Article and Find Full Text PDF

In silico screening and immunogenic features of putative tick cement protein PA107 from Ixodes ricinus tick.

Exp Appl Acarol

January 2025

Group for Medical Entomology, Centre of Excellence for Food- and Vector-Borne Zoonoses, Institute for Medical Research, National Institute of Republic of Serbia, University of Belgrade, Belgrade, Serbia.

Tick salivary proteins are crucial for efficient and successful tick feeding. Most of them are still uncharacterized, especially those involved in the formation of tick cement. Tick salivary protein PA107 is a putative cement protein, which is transcribed in salivary glands during the initial phase of tick feeding.

View Article and Find Full Text PDF

Objectives: We evaluated the performance of a novel flow cell morphology analyzer AUTION EYE AI-4510 for counting particles in urine.

Methods: Analytical performance was assessed according to the EFLM European Urinalysis Guideline 2023. Trueness was compared by analyzing 1.

View Article and Find Full Text PDF

Biomechanics of the bobsleigh push phase.

J Sports Sci

January 2025

Institute of Biomechanics and Orthopaedics, German Sport University Cologne, Cologne, Germany.

The purpose of this work was to provide a fundamental, in-depth analysis of kinematics and kinetics of the bobsleigh push phase to establish a basis for performance analysis and enhancement. Fifteen elite male athletes performed maximal effort push starts, while ground reaction forces (GRF) and 3D marker trajectories were simultaneously recorded for ground contacts of different sub-sections of the push phase (start acceleration phase: first and second ground contact after the initial push-off from the start block, acceleration phase: 10 m and high-velocity phase: 30 m). To obtain a comprehensive view of the push phase, whole-body kinematics as well as joint kinetics were analysed and compared across the push phase.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!